Stryker IVEA 600S Assembly Instructions Download Page 1

READ THIS FIRST! 

IVEA 600S Assembly Instructions 

Tools required: None    Estimated time: 5 minutes

 

IMPORTANT: Casters include left and right leg casters, marked L and R, with blue and yellow pedals, 

and two identical nose casters with overlapping blue pedals. If installed improperly, the SureTrek

®

 

tracking system will not work. 

1)

 Remove all 4 casters and the IV Hooks box from the carton. Remove the IVEA from its plas-

tic sleeve and position the collapsed device so that the undersides of the legs face upward.   

2) 

Locate the leg caster marked " 

L

" on the hex 

surface. Engage the brake by depressing the 

blue pedal. Partially insert the caster stem into 

the end socket of the 

left 

leg, which is on the 

same side as the oxygen tank hoop. Rotate the 

caster until the notch in the hex is positioned 

exactly as shown, directly in line with its mating 

"key." Push caster fully into place, making sure 

hex engages its socket. 

Hex will not engage 

and SureTrek

®

 tracking system will not work 

if notch is improperly oriented! 

 

3) 

Repeat Step 2 with the right leg caster 

(marked“

R

"), making certain the notch is orient-

ed correctly and hex is fully engaged. 

4) 

Engage the brake of a nose caster by de-

pressing the blue pedal marked “ON.” Insert the 

caster stem into a socket in the IVEA nose. Ro-

tate the caster until the hex sides align with the 

socket, then push fully into place. 

Note that 

nose casters do not have orientation notches. 

 

5) 

Repeat Step 4 with the second nose caster. 

6) 

Release the brakes on all four casters by pull-

ing up on the blue pedals. 

Deploy the IVEA by 

carefully following the instructions on the re-

verse side of this sheet. 

When all 4 casters are 

correctly installed, the deployed IVEA will rest 

stably on the floor. If not stable, check that all 

caster hexes are fully engaged in their mating 

sockets.

 

INCORRECT: 

Hex not fully engaged 

CORRECT: 

Hex fully engaged 

Hex notch must be 

oriented as shown 

Key 

IMPORTANT: Inspect device for any damage that may have occurred during shipment.  If damage is 

found, DO NOT put device into service and call Stryker Customer Service at: 1-800-327-0770. 