READ THIS FIRST!
IVEA 600S Assembly Instructions
Tools required: None Estimated time: 5 minutes
IMPORTANT: Casters include left and right leg casters, marked L and R, with blue and yellow pedals,
and two identical nose casters with overlapping blue pedals. If installed improperly, the SureTrek
®
tracking system will not work.
1)
Remove all 4 casters and the IV Hooks box from the carton. Remove the IVEA from its plas-
tic sleeve and position the collapsed device so that the undersides of the legs face upward.
2)
Locate the leg caster marked "
L
" on the hex
surface. Engage the brake by depressing the
blue pedal. Partially insert the caster stem into
the end socket of the
left
leg, which is on the
same side as the oxygen tank hoop. Rotate the
caster until the notch in the hex is positioned
exactly as shown, directly in line with its mating
"key." Push caster fully into place, making sure
hex engages its socket.
Hex will not engage
and SureTrek
®
tracking system will not work
if notch is improperly oriented!
3)
Repeat Step 2 with the right leg caster
(marked“
R
"), making certain the notch is orient-
ed correctly and hex is fully engaged.
4)
Engage the brake of a nose caster by de-
pressing the blue pedal marked “ON.” Insert the
caster stem into a socket in the IVEA nose. Ro-
tate the caster until the hex sides align with the
socket, then push fully into place.
Note that
nose casters do not have orientation notches.
5)
Repeat Step 4 with the second nose caster.
6)
Release the brakes on all four casters by pull-
ing up on the blue pedals.
Deploy the IVEA by
carefully following the instructions on the re-
verse side of this sheet.
When all 4 casters are
correctly installed, the deployed IVEA will rest
stably on the floor. If not stable, check that all
caster hexes are fully engaged in their mating
sockets.
INCORRECT:
Hex not fully engaged
CORRECT:
Hex fully engaged
Hex notch must be
oriented as shown
Key
IMPORTANT: Inspect device for any damage that may have occurred during shipment. If damage is
found, DO NOT put device into service and call Stryker Customer Service at: 1-800-327-0770.