T
Trro
ou
ub
blle
essh
ho
oo
ottiin
ng
g
P
Prro
ob
blle
em
m
S
Sccrre
ee
en
n
C
Ca
au
usse
e
R
Re
ecco
om
mm
me
en
nd
de
ed
d a
accttiio
on
n
Kinked air hoses
Air hose is bent or an
obstruction in the hose may
cause air flow to be
compromised
T
Trru
uT
Tu
urrn
n was attempted but
you have a standard Model
(bottom cover is orange).
1. Make sure the hoses are
straight and air flow is not
obstructed.
2. Press the action button
next to the Alarm off icon
to reset the pump.
Contact sales support for
T
Trru
uT
Tu
urrn
n option
Missed connection with air
hoses
Air hoses is not connected
to the pump or the support
surface
1. Make sure that the hoses
are seated all the way
onto the hose ports on
the pump or the support
surface.
2. Press the action button
next to the Alarm off icon
to reset the pump.
Power loss, product does
not turn on
Power cord not seated,
power cord unplugged from
outlet, or possible internal
damage
Make sure the power cord is
plugged into the product and
the outlet.
Power loss, while T
Trru
uT
Tu
urrn
n
is in use
Unintentional or intentional
power loss. Turn bladder is
inflated.
Pull the hose from the pump
or the support surface to
activate CPR. The bladder
deflates and returns the
support surface to a flat
position.
Button not responsive
29:59
Lock function active
Press and hold the action
button next to the locked icon
for 2 seconds to turn off the
lock.
T
Trru
uT
Tu
urrn
n not shown on
screen
Graphical user interface is
for the standard model, not
connected to a premium
IIsso
oT
To
ou
urr bottom cover with
anchor straps
Check for kinked hoses.
Verify that you have premium
bottom cover (black).
Contact sales support for
T
Trru
uT
Tu
urrn
n option
T
Trru
uT
Tu
urrn
n shown on screen
Graphical user interface
shows T
Trru
uT
Tu
urrn
n but you
have a standard Model
(bottom cover is orange).
Turn off the pump to reset
the pump features. Make
sure the air hose is
connected before you turn
the pump on.
EN
16
2874-009-001 Rev B.0