Stryker Adel 4700 Operation Manual Download Page 15

14

Operating Bed Features

REMOVING THE HEAD BOARD

To remove the head board, lift it straight up and off the bed.  To replace the head board, align the plastic inserts
on the bottom of the head board with the pins on the bed and lower the head board until it completely engages
the pins.

USING THE LABOR GRIPS

To position the labor grips, grasp the handle and pull the grip out and up above the mattress until the mecha-
nism locks into position.  To lower the grips, pull the lever to release the grip and, as it begins to rotate, release
the lever and lower the grip to the stored position.

ADJUSTABLE SEAT DEPTH (OPTIONAL)

To accommodate patients of different sizes, the depth of the seat can be adjusted up to 4 inches (from 12.5
to 8.5 inches).

S

The Fowler (back rest) must be raised to a minimum of 35

_

 before the seat depth can be changed.  Once

the Fowler (back rest) is elevated to 35

_

, the “ENABLED” LED on the outside of the siderail will light and

the seat depth can be changed.  When the Fowler (back rest) is lowered (either electrically or using the
emergency CPR release), the bed will automatically adjust the seat depth to 12.5”.

S

To reduce seat depth, depress the 

 button on the siderail or the hand pendant to move the Fowler

(back rest) toward the foot end of the bed.  Release the button when the desired seat depth is attained.

S

To increase seat depth, press the 

 button to move the Fowler (back rest) toward the head end

of the bed.

NURSE CALL BATTERY

To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head
end of the bed to the off position.  The switch is identified by the label shown below.  If the switch is not posi-
tioned as shown below and the bed power cord and pendant cord are unplugged, the life of the back−up bat-
tery will be significantly reduced.
If the siderail lights LED (located on the outside of both siderails) is flashing, the Nurse Call battery needs
to be  replaced.  The battery is located on the patient’s right side at the head end of the bed.  No tools are
required to replace the battery.  Unplug the bed power cord from the wall socket and replace the battery.  After
replacing the battery, verify the siderail lights LED is no longer flashing and operates normally when the differ-
ent light settings are selected.

Summary of Contents for Adel 4700

Page 1: ...Adelr Maternity Bed Model 4700 5012 OPERATIONS MANUAL For Parts or Technical Assistance 1 800 327 0770 ...

Page 2: ...3 Using the Labor Grips 14 Adjustable Seat Depth 14 Nurse Call Battery 14 Night Light Usage 15 Operating the Foot Section 15 Lumbar and Perineal Pillows 15 Patient Positioning Birthing Bed Positioning 16 Birthing Chair Positioning With Labor Bar 16 Delivery Table Positioning 16 Trendelenburg Positioning 16 Pelvic Tilt 16 Using Accessories Using Leg Supports 17 Using the Labor Bar 17 Using the Arm ...

Page 3: ... Caster Size 8 Standard 6 Optional Stryker reserves the right to change specifications without notice WARNING CAUTION NOTE DEFINITION The words WARNING CAUTION and NOTE carry special meanings and should be carefully reviewed WARNING The personal safety of the patient or user may be involved Disregarding this information could result in injury to the patient or user CAUTION These instructions point...

Page 4: ...nal safety precaution in that means are provided for the connection of the EQUIPMENT to the protective earth con ductor in the fixed wiring of the installation in such a way that ACCESSIBLE METAL PARTS cannot become live in the event of a failure of the BASIC INSULATION Mode of Operation Continuous IPX4 Protection from liquid splash Dangerous Voltage Symbol Protective Earth Terminal Potential Equa...

Page 5: ...R release is for emergency use only When activating the instant CPR all persons and equipment must be removed from the area below and around the Fowler back rest section of the bed or serious personal injury or damage to the equipment could occur S When large spills occur in the area of the circuit boards 110 volt cables and motors immediately unplug the bed power cord from the wall socket Remove ...

Page 6: ...iderail control option only S Plug the interface cable into the 37 pin connector in the litter frame at the head end of the bed and into the Patient Station Head Wall Docker Station or equivalent whichever applies S Run through each function on the siderail control panels to ensure that each function is working properly NOTE To prevent a low battery condition when the bed is not plugged in positio...

Page 7: ...6 Bed Illustration WARNING Potential pinch points ...

Page 8: ...n when it is raised USING THE SIDERAIL CONTROL PANEL LIGHTS The bed is equipped with lights to illuminate the head end siderail control panels and the red nurse call switches Four settings are available for the control panel lights Push once to turn on the lights at low intensi ty again to change to high intensity a third time to leave just the nurse call light on and a fourth time to turn all lig...

Page 9: ...n the room light on Push again to turn off 2 Push to turn the bed overhead light on Push again to turn off 1 Push for more support to the patient s lower back 2 Push for less support to the patient s lower back 3 Push for more support to the patient s seat section 4 Push for less support to the patient s seat section 5 Push to change the channel on the television 6 Push to change the channel on th...

Page 10: ...s function is also used for pelvic tilt 2 Push to toggle siderail lights to different settings A LOW B HIGH C NURSE CALL ONLY D OFF 3 Push to raise head end of bed and or return bed to level 4 Push to lock out all bed motion Push again to unlock LED will light when bed motion is locked 5 Push to activate Nurse Call 1 Push to decrease seat depth 2 Push to increase seat depth NOTE The Fowler back re...

Page 11: ...10 Communication Pendant Operation Guide ...

Page 12: ...11 Motion Pendant Operation Guide ...

Page 13: ...12 Motion Pendant with Nurse Call Operation Guide ...

Page 14: ... crank stored in clips on the head end of the bed Center the notch in the end of the crank on the roll pin on the bed motor assembly or the Fowler motor assembly and crank the bed to the desired position ACTIVATING INSTANT CPR The CPR release lever is located at the head section on both sides of the bed To activate the CPR release grasp the lever and squeeze tightly The Fowler back rest will lower...

Page 15: ...g the emergency CPR release the bed will automatically adjust the seat depth to 12 5 S To reduce seat depth depress the button on the siderail or the hand pendant to move the Fowler back rest toward the foot end of the bed Release the button when the desired seat depth is attained S To increase seat depth press the button to move the Fowler back rest toward the head end of the bed NURSE CALL BATTE...

Page 16: ...ioned 6 below the top of the seat mattress to accommodate different sized patients when the foot rest or calf supports are being used S To store the foot rests loosen the ball screws on the self storing calf supports if used Ensure the foot rests are pointing toward the end of the bed Pull the lever marked PULL located at the bottom of the foot rest support and the foot rest will glide under the f...

Page 17: ... TABLE POSITIONING S Slide the patient down to the perineal edge S Lower the foot section to its lowest position S Position the attached calf supports and place the patient s legs in the supports TRENDELENBURG POSITIONING Lower the Fowler back rest by pressing the control on the siderail or by squeezing the CPR lever Press the Trendelenburg button on the siderail or the control pendant to lower th...

Page 18: ...ar insert it into the holes on each side of the seat section under the mattress at the perineal edge The labor bar can be used by the patient to aid with various birthing positions such as squatting or kneeling USING THE ARM BOARD The arm board is held in place between the mattress and the frame The arm board is provided with a remov able mattress The weight of the patient retains the arm board in...

Page 19: ... pole pull up on the telescoping portion A until it locks into place at its fully raised position 3 For a higher I V pole pull up on section B Release section B at any desired height and it will lock into place 4 Rotate the I V hangers C to the desired position and hang the I V bags 5 To lower the I V pole push up on the red portion of grip D while holding onto section B until it lowers Turn latch...

Page 20: ...2 ounce gallon Matar Phenolic Huntington Laboratories 1 2 ounce gallon Omega Quaternary Airwick Professional Products Division 1 2 ounce gallon Quanto Quaternary Huntington Laboratories 1 ounce gallon Sanikleen Quaternary West Chemical Products 2 ounces gallon Sanimaster II Quaternary Service Master 1 ounce gallon Vesphene Phenolic Vestal Laboratories 1 1 4 ounce gallon Wexcide Phenolic Wexford La...

Page 21: ... Trendelenburg operates properly I V pole intact and operating properly No rips or cracks in mattress cover Lubricate where required Replace Nurse Call 9V battery annually Power cord not frayed No cables worn or pinched All electrical connections tight All grounds secure to the frame Ground impedance not more than 100 milliohms Current leakage not more than 100 microamps Serial No NOTE Preventativ...

Page 22: ... maintenance manual for each device This statement constitutes Stryker s entire warranty with respect to the aforesaid equipment STRYKER MAKES NO OTHER WARRANTY OR REPRESENTATION EITHER EXPRESSED OR IMPLIED EXCEPT AS SET FORTH HEREIN THERE IS NO WARRANTY OF MERCHANTABILITY AND THERE ARE NO WARRANTIES OF FITNESS FOR ANY PARTICULAR PURPOSE IN NO EVENT SHALL STRYKER BE LIABLE HEREUNDER FOR INCIDENTAL...

Page 23: ... without approval from the Stryker Customer Service Department An autho rization number will be provided which must be printed on the returned merchandise Stryker reserves the right to charge shipping and restocking fees on returned items SPECIAL MODIFIED OR DISCONTINUED ITEMS NOT SUBJECT TO RETURN Damaged Merchandise ICC Regulations require that claims for damaged merchandise must be made with th...

Page 24: ...V B European Representative Stryker EMEA RA QA Director Stryker France ZAC Satolas Green Pusignan Av De Satolas Green 69881 MEYZIEU Cedex France 6300 S Sprinkle Road Kalamazoo MI 49001 9799 800 327 0770 www strykermedical com ...

Reviews: