14
Guidance and manufacturer's declaration - electromagnetic immunity
The Cast Cutter REF 941 is intended for use in the electromagnetic environment specified below. The customer or the user of the
Cast Cutter REF 941should assure that it is used in such an environment.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
IEC 60601 test level
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
Compliance level
3 Vrms
3 V/m
Electromagnetic environment - guidance
Portable and mobile RF communications equipment should
be used no closer to any part of the Cast Cutter REF
941, including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d=1.67√
P
d=1.67√
P
80 MHz to 800 MHz
d=2.33√
P
800 MHz to 2.5 GHz
Where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d
is the recommended separation
distance in meters (m)
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1: At 80 MHz and 800MHz the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.