The adjustable cot load height with jog function allows you to preset the height of the load wheels to meet the ambulance
deck height up to 36 in. The jog function assists operators when loading on an incline by jogging past the preset load height.
You can set the cot load height from 26 in. to 36 in. (66 cm to 91 cm) as measured from the ground to the bottom of the load
wheel.
To set the cot load height:
1. Locate the sensor housing on the patient right side of the cot.
2. Using a T27 Torx driver, remove the sensor housing cover by loosening the two screws (one on each end).
3. Adjust the left height sensor only inside the sensor housing: (Figure 11):
a. Move the sensor toward the head end to increase the set load height or move the sensor toward the foot end to
decrease the set load height.
b. Press the retract (-) button to lower the cot to its lowest position, then press the extend (+) button to raise the cot to
its set load height position.
c. Measure the cot height from the bottom of the load wheels to the floor.
N
No
otte
e -- Add an additional 1/2 in. (1,3 cm) to your deck height measurement to allow for variations in patient weight or
with other equipment that may be added to the cot.
d. Repeat steps 3a and 3b until you reach the desired cot load height.
4. Secure the height sensor cables. All cables should lie flat inside the housing between the sensors.
5. Using a T27 Torx driver, reinstall the two screws (removed in step 2) to replace the sensor housing cover.
F
Fiig
gu
urre
e 1
11
1 –– A
Ad
djju
usstt lle
efftt sse
en
nsso
orr h
he
eiig
gh
htt
C
Ch
he
ecckkiin
ng
g tth
he
e b
ba
atttte
erryy p
po
ow
we
err lle
evve
ell
Use the cot battery LED indictor to check the S
SM
MR
RT
T Pak power level. A charged S
SM
MR
RT
T Pak, in working condition, provides
up to 25 calls with a 250 pound patient (actual results may vary). The 24 VDC P
Po
ow
we
err--P
PR
RO
O system and the S
SM
MR
RT
T Pak is
rated for 2.4 amp-hours of electric energy.
W
WA
AR
RN
NIIN
NG
G -- Do not remove the battery when the cot is active.
C
CA
AU
UT
TIIO
ON
N -- Always charge the battery before you place the product into service. An uncharged or depleted battery may
cause poor product performance.
To check the battery power level, press the retract (-) button on the cot control switch to activate the cot battery LED
indicator. The cot battery LED indicator is located at the foot end control enclosure (shown as a battery symbol).
EN
22
6516-209-001 Rev D.0
Summary of Contents for 6516
Page 2: ......
Page 62: ...6 65 50 06 60 00 00 01 10 00 01 16 6 Rev AA A A 650600010130 2 EN 58 6516 209 001 Rev D 0 ...
Page 63: ...6 65 50 00 0 0 00 02 2 0 02 28 8 Rev A A A 6500 001 160 1 6516 209 001 Rev D 0 59 EN ...
Page 64: ...6 65 50 00 0 1 10 02 2 0 01 15 5 Rev AA A A 6500 102 014 1 EN 60 6516 209 001 Rev D 0 ...
Page 66: ...6 65 50 00 0 0 00 02 2 1 15 59 9 Rev AA A A 6500 002 159 1 EN 62 6516 209 001 Rev D 0 ...
Page 67: ...6 65 50 00 0 2 20 01 1 1 14 48 8 Rev AA A A 6500 201 248 1 6516 209 001 Rev D 0 63 EN ...
Page 68: ...6 65 50 00 0 2 20 01 1 0 01 10 0 Rev AA A A 6500 201 140 1 EN 64 6516 209 001 Rev D 0 ...