A charged battery, in working condition, provides at least 14 lifts with a 250 lb (113.4 kg) patient (actual results may vary).
The charger LED indicator bars show the battery power level.
F
Fiig
gu
urre
e 9
9 –– B
Ba
atttte
erryy p
po
ow
we
err lle
evve
ell
S
Stta
attu
uss
B
Ba
atttte
erryy L
LE
ED
D iin
nd
diicca
atto
orr
Discharging
Four LED bars = 76-100% charge
Three LED bars = 51-75% charge
Two LED bars = 26-50% charge
One LED bar = 15-25% charge
Low battery
<15% charge
One LED blinks five times, repeated two to three times
Charging
LED indicating current charge percentage is blinking
Error
Press LED indicator button and outermost LEDs blink five
times, repeated three times
See
Troubleshooting
in the Maintenance Manual
N
No
otte
e
•
Use only Stryker approved batteries.
•
Lock the cot into the powered cot fastener to automatically charge the battery. There are no cables or connectors
required to charge.
T
Trra
an
nssffe
errrriin
ng
g tth
he
e p
pa
attiie
en
ntt tto
o tth
he
e cco
ott
W
WA
AR
RN
NIIN
NG
G
•
Always use all restraint straps to secure the patient on the product. An unrestrained patient may fall from the product and
be injured.
6507-009-001 Rev AB.3