Stryker 6400-099-000 Instructions For Use Manual Download Page 8

8

www.stryker.com

EN

6400-099-700 Rev-AA

Instructions (continued)

To Operate the Handpiece

NOTE: 

See the instructions for use supplied with the appropriate 

console for additional information about handpiece operation.
1. Install an accessory into the attachment.

NOTE: 

See the instructions for use supplied with the attachment to 

properly install an appropriate accessory.
2. Plug one end of the cord into the handpiece cord receptacle 

and the other end of the cord into the appropriate console cord 
receptacle.

3. Use the console to program the operational settings of the 

handpiece as required.

NOTES:

•  When using the console, the Forward and Reverse buttons on 

the Universal Driver Screen do not function. Use the handpiece 
triggers or a footswitch to select the handpiece direction.

•  When using the console, the Oscillate button will initiate a 

turn-based oscillate mode that defaults to a one-turn oscillation.  
Use the console touchscreen to select more turns as required.

•  Depress both triggers together to operate the handpiece in the 

oscillate mode. 

•  Pressure on a trigger will also control the speed of the 

handpiece.

4. Use either the handpiece triggers or a footswitch to operate the 

handpiece.

NOTE: 

Use the console touchscreen to assign different functions 

(such as the oscillate mode) to the footswitch pedals as required.

To Remove an Attachment

1. Fully depress the release button.
2. Firmly grasp and pull the attachment from the attachment 

connector.

Final Disassembly

1. Unplug the cord from the console and from the handpiece.
2. Remove attachment as described above.

Summary of Contents for 6400-099-000

Page 1: ...2018 10 6400 099 700 Rev AA www stryker com Instructions For Use RemB Electric Universal Driver 6400 099 000 ENGLISH EN...

Page 2: ...t product damage A NOTE supplements and or clarifies procedural information For additional information especially safety information or in service training contact your Stryker sales representative or...

Page 3: ...fore use See the Specifications section Upon initial receipt and before each use clean and sterilize the equipment as indicated See the care instructions manual supplied with the equipment Upon initia...

Page 4: ...acement and or bone or tissue damage If wobbling occurs see the Troubleshooting section When using the Universal Driver always use Stryker attachments with two J slots The Universal Driver will not fu...

Page 5: ...fety Certification standards listed in the Specifications section of the instructions for use supplied with the console D Attachment E Reverse Trigger Press this pressure sensitive trigger to obtain v...

Page 6: ...ctional FORWARD REVERSE Fully extend the switch to unlock both triggers Both triggers are functional Definitions The symbols located on the equipment and or labeling are defined in this section or in...

Page 7: ...e is idle when passing the handpiece to another person or before inserting or removing a cutting accessory or attachment DO NOT press the pedal of a connected footswitch while the function switch is s...

Page 8: ...en do not function Use the handpiece triggers or a footswitch to select the handpiece direction When using the console the Oscillate button will initiate a turn based oscillate mode that defaults to a...

Page 9: ...trical malfunction exists Return the handpiece for repair The handpiece will not operate The function switch is in the SAFETY position Set the function switch to the FORWARD or FORWARD REVERSE positio...

Page 10: ...rference is experienced during installation The release button mechanism needs lubrication See the care instructions manual supplied with the equipment Sporadic electrical interference is experienced...

Page 11: ...Cycle Rest Between Cycles Non continuous Operation 20 seconds on 20 seconds off 6 times 45 minutes Power Supply Stryker CORE 2 CORE or TPS Console 40 V Direct Current Equipment Type Type BF Applied Pa...

Page 12: ...700 Rev AA www stryker com Stryker Instruments 4100 E Milham Kalamazoo Michigan USA 49001 1 269 323 7700 1 800 253 3210 ES DE FR IT NL 6400 099 713 JA ZH KO 6400 099 720 SV DA FI PT NO 6400 099 730 PL...

Reviews: