Operation Guide
OPERATING THE ENHANCED CLEARANCE HEAD PIECE
To operate the enhanced clearance head piece, grasp either handle under the
head section and squeeze. Handle (A) releases the latch and rotates the
head piece on axis ”A”. Handle (B) releases the other latch and rotates
the head section on axis ”B”. For ease of operation, it is recom-
mended that only one latch be released at a time.
WARNING
The weight of the patient’s head is resting on the head piece and must be supported by the operator when
the latches are released and the head piece is being positioned. Failure to adequately support the head piece
while positioning the head could result in patient injury.
CAUTION
The dual articulating headpiece is designed to provide precision surgical positioning. Be sure to treat it with
care. The unit should be routinely checked to ensure optimal performance. In the event of any impact or
overload of the headpiece, be sure the unit is working properly and supports the intended load. Verify the
adjustment gears lock and release properly and, if necessary, refer to the adjustment procedure in the mainte-
nance manual.
USING THE OPTIONAL INFLATABLE HEAD SUPPORT CUSHION
The optional inflatable head support cushion has two internal air blad-
ders. Squeeze the bulb (A) to inflate the bladders and provide more sta-
bility for the patient’s head. Press the release valve (B) to deflate the
bladders.