Instructions
To Install Non-sterile Battery
WARNING:
DO NOT use the aseptic
housing, transfer shield, or non-sterile
battery unless they are dry.
CAUTION:
DO NOT allow the battery contact
terminals to touch metal objects.
NOTE:
Ensure more than one sterilized
battery protector kit is available in case a
second battery installation is necessary.
Sterile Assistant Action
1. Open the lid (Figure 1).
Figure 1 – Open Lid
2. With the aseptic housing chamber open,
place the transfer shield onto the aseptic
housing and over its chamber (Figure 2).
Figure 2 – Place Transfer Shield
3. Present the chamber to the circulating
assistant.