At a minimum, check all items listed during annual preventive maintenance for all Stryker Medical products. You may need
to perform preventive maintenance checks more frequently based on your level of product usage.
Remove product from service before performing preventive maintenance.
N
No
otte
e -- Perform support surface care before inspection, if applicable.
Inspect the following items:
Zipper and covers (top and bottom) are free of tears, cracks, cuts, holes, or other openings
Internal components for signs of staining from fluid ingress or contamination by fully unzipping the covers
Labels for legibility, proper adherence, and integrity
Foam has not degraded or come apart
Product serial number:
Completed by:
Date:
A
Acccce
esssso
orriie
ess a
an
nd
d p
pa
arrttss
These parts are currently available for purchase. Call Stryker Customer Service: 1-800-327-0770 for availability and pricing.
N
Na
am
me
e
N
Nu
um
mb
be
err
Cover assembly 30”
1703-034-335
Cover assembly 30”
1704-034-335
Cover assembly 26”
1704-034-635
C
Co
ovve
err rre
ep
plla
acce
em
me
en
ntt
W
WA
AR
RN
NIIN
NG
G -- Do not wash the internal components of this support surface. Discard the support surface if a contamination is
found inside.
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
None
P
Prro
occe
ed
du
urre
e::
1. Unzip the cover.
2. Fold the top of the cover to the patient’s right side and then remove the foam crib assembly from the stretcher and set
aside.
3. Discard the old cover.
4. Place the new cover, unzipped and open, on the frame with the grey bottom cover on the litter and the top cover folded
over the patient’s right side of the stretcher.
5. Carefully place the foam crib assembly on top of the bottom part of the cover to make sure that the foam crib is Align
with the cover.
6. Fold the top cover over the top of the foam crib assembly. Make sure that the top cover is aligned with the foam crib
assembly.
7. Zip the cover to close.
EN
10
1704-009-001 Rev F.0