G -- Always clean and disinfect the rover upon initial receipt and before each use. Failure to comply may
cause infection and result in patient or healthcare staff injury.
C
CA
AU
UT
TIIO
ON
N -- Do not immerse any system component in liquid. Do not allow liquids or moisture to enter any electrical
connection.
C
CA
AU
UT
TIIO
ON
N -- Do not sterilize any system component.
C
CA
AU
UT
TIIO
ON
N -- Do not use solvents, lubricants, or other chemicals, including glutaraldehyde or similar chemical cleaners,
unless otherwise specified. See
C
CA
AU
UT
TIIO
ON
N -- Do not use unapproved disinfectants. Failure to comply may cause system damage.
R
Re
ecco
om
mm
me
en
nd
de
ed
d E
Eq
qu
uiip
pm
me
en
ntt a
an
nd
d M
Ma
atte
erriia
allss
•
PPE as recommended by the disinfectant supplier
•
Soft, lint-free cloth
•
United States Environmental Protection Agency (US EPA) registered disinfectant with a claim for activity against
Hepatitis B. The following disinfectants have been validated for use on the exterior surfaces of the Stryker Neptune 3
Waste Management System: Sodium Hypochlorite Based - Clorox® Clean-Up® Disinfectant Cleaner with Bleach (EPA
Reg. #67619-17), Quaternary Ammonium Based - CaviCide® (EPA Reg. #46781-6)
5
5..1
1..1
1.. T
To
o W
Wiip
pe
e D
Do
ow
wn
n tth
he
e R
Ro
ovve
err
1. Wipe all external surfaces of the rover with a soft, lint-free cloth moistened with a non-abrasive, hospital disinfectant
prepared to the manufacturer’s instructions. Pay particular attention to highlighted critical areas such as the work
surface, handle, main control panel, communication window, and manifold receptacles (Figure 39).
F
Fiig
gu
urre
e 3
39
9 –– W
Wiip
pe
e D
Do
ow
wn
n tth
he
e R
Ro
ovve
err
2. Upon the removal of visible, gross soil, use a clean cloth moistened with disinfectant and wipe all surfaces. Surfaces
must remain visibly wet at room temperature for at least the minimum time specified in the instructions for use supplied
by disinfectant manufacturer.
3. Remove any excess disinfectant solution using a soft, lint-free cloth moistened with water if required by the instructions
supplied by the disinfectant manufacturer.
EN
50
700000969828 Rev-AA
DRAFT