Service Information
HYDRAULIC SYSTEM TROUBLESHOOTING
NOTE
Be sure the pedal linkage has been adjusted properly before beginning service on the jacks (see page 6
or page 7).
PROBLEM/SYMPTOM
SOLUTION
Jack will not raise to full height.
Add hydraulic fluid (see p.11). Check for leaks.
Jack will not hold in raised position.
Close the needle valve completely. If the jack
holds, replace valve #1 (see p. 14). If the jack
does not hold, replace valve #2 (see p. 15).
Jack will not pump up and the jack actuator rod
does not
move.
Close the needle valve. If the jack will now pump
up, replace valve #1. If the jack still will not pump
up after closing the needle valve, replace valve #3
(see p. 15).
Jack will not pump up but the jack actuator rod
does
move when the pump pedal is activated.
Replace valve #2 (see p. 15).
Jack will not pump up and the jack actuator rod
may or may not move.
Remove excess air (vacuum) in system (see p.
11).
Contact Stryker technical service at 1–800–327–0770 for further assistance.
Summary of Contents for 1068
Page 10: ...5 4 Notes ...
Page 21: ...1068 1 10 Side Control Base Assembly 16 ...
Page 23: ...1068 1 10 Side Control Base Assembly 18 ...
Page 27: ...Notes 20 2 ...
Page 28: ...1068 2 10 End Control Base Assembly 21 ...
Page 30: ...1068 2 10 End Control Base Assembly 23 ...
Page 42: ......
Page 46: ...1068 30 10 Litter Assembly 34 HEAD END ...
Page 58: ...1068 232 10 Enhanced Clearance Fowler Subass y 43 1 ...
Page 60: ...1068 32 10 Crank Fowler Subassembly 44 ...
Page 86: ...Notes 62 2 ...