issue V8.3E from
01.07.2019
13
4.4.1
Chemical-thermal disinfection
In the case of suspected contamination, the used vacuum lines should immediately be submitted to chemical-
thermal disinfection. With it, the vacuum lines will be submitted to a validated process of exterior and interior
cleaning/disinfection (for informations please ask the manufacturer).
4.5
ECG plug-in cable
The plug-in connector of the ECG plug-in cable at the support arm is connected to the ECG standard connector.
Other ECG devices are connected via adapters.
4.6
Maintenance
The devices should be checked (STK) according to the latest instructions given by the manufacturer at intervals of
2 years or after repair (or after intervention in the device). The check should be carried out by persons authorized
by Strässle & Co Medizintechnik GmbH. The exchange of fuses on the power supply should be effected only by
qualified personnel. Disconnection from the mains is compulsory.
The power cable must be disconnected!
No
other fuses than 2 x 0,315AT are allowed.
No service or maintenance should be performed while the unit is being used on a patient.
4.7
Specifications
Medical product class I (93/42/EWG)
Supply voltage
230V~ 50Hz
Rated consumption
22VA
Vacuum range
60hPa - 200hPa
Protection class
I
Protection type
IP 40
Degree of protection BF
(supporting arm BF or CF defibrillation protected)
Dimensions
162x190x75 (w x l x h) vacuum unit only
4.8
Scope of delivery (depending on model version)
Power cable
(1000-099) -
removable part
Vacuum unit with control unit
(3000-205)
Holder for control unit
(2000-237) -
accessory
Distributor
(3000-177)
Support arm
(3000-178)
Clips for cable adjustment
(1000-596-2) -
accessories
Electrode set 6x1m / 4xEXT 1,3m
(4000-005) -
removable parts
(alternatively electrodes 10x1m or 10x1,3m) (4000-001 or 4000-003) -
removable parts
Electrode spray
(1000-158) -
accessory
Velcro fastening strips
(2000-463) -
accessories
optionally with mounting kit (various mounting kits possible) -
accessories
4.9
Warranty
Strässle & Co. Medizintechnik GmbH
grants warranty according as stated in the terms for sales, delivery, and
payment. Wearing parts and disposable material are not included in warranty.
Warranty will lapse in case of
damage caused by improper operation and undue usage.
inaccurate assembly, actions taken by unauthorized personnel, or use of accessories, disposable material or
replacement parts which are not original parts of Strässle & Co. Medizintechnik GmbH.
changes, extensions, repairs or actions whatsoever effected by persons not authorised by the manufacturer.
the electrical installation of the location where the device is connected not complying with the requirements
according to VDE 0107.
using the device disregarding the operating instructions.
Even after warranty has expired, original parts and accessories of the manufacturer should be used. Only thereby safe
and proper operation can be ensured, as these products are continually improved and optimized.