19 620 02 0117
66
Technical
Specifications
For the technical specifications of the handpieces, please refer to the operating
manual for your particular handpiece.
Equipment safety (“essential performance“) according to IEC 60601-1, 3rd edition:
Applied acoustic energy does not exceed the specified limit of 6.5 bar with a tolerance
of 10%.
The applied underpressure with the suction cup is limited to -500 mbar.
NOTE
When the medical product is distributed to third parties, the following must be
observed:
– The complete device documentation must be delivered together with the
medical product.
– The medical product may only be exported to a foreign country when the
medical product and the corresponding indications are allowed there.
7.2
Type plate D-ACTOR
®
200
Fig. 7-1 Type plate voltage supply 220 - 230 VAC
Fig. 7-1 Type plate voltage supply 115 VAC
7.3
Conformity with directives
This medical product bears the CE mark in accordance with the
Medical Device Directive (MDD) 93/42/EEC