background image

 

Instructions for use
CLICKLINE Instruments 10 mm

en

Summary of Contents for CLICKLINE Instruments 10 mm

Page 1: ...Instructions for use CLICKLINE Instruments 10 mm en...

Page 2: ...illustrations product descriptions and texts are the intellectual property of KARL STORZ SE Co KG Their use and reproduction by third parties require the express approval of KARL STORZ SE Co KG All ri...

Page 3: ...the field of vision 9 3 6 Peak voltage output 10 3 7 Electrical contacts 10 3 8 HF applied parts 10 3 9 Tissue deposits due to coagulation 10 3 10 HF electrodes 10 3 11 Risk of explosion and fire 10 3...

Page 4: ...tructions for use CLICKLINE Instruments 10 mm TBF107_EN_V1 0_10 2021_IFU_CE MDR 4 9 Accessories and spare parts 21 9 1 Recommended accessories 21 9 2 Spare parts 22 10 Appendix 23 10 1 Appendix Scope...

Page 5: ...or the product may be damaged Read the instructions for use of the compatible products carefully and follow all the safety notes and warnings Read the reprocessing instructions of the compatible prod...

Page 6: ...warnings and safety notes in the instructions for use must be observed The warnings use the following levels of danger WARNING WARNING Designates a possible imminent risk If this is not avoided it co...

Page 7: ...hs are surgically invasive and meant for short term use Forceps insert Forceps inserts are used for grasping preparing and dissecting tissue in minimally invasive procedures Forceps inserts are surgic...

Page 8: ...devices must not be used for procedures in direct contact with the central nervous system CNS and central cardiovascular system Beyond that there are no contraindications for the use of the medical d...

Page 9: ...ning prions Prions lead to infection with Creutzfeldt Jakob disease If Creutzfeldt Jakob disease has been diagnosed or is suspected Do not continue to use the product Dispose of the product properly 3...

Page 10: ...s users and third parties if there is unintentional contact Do not place HF applied parts on the patient Do not let HF applied parts come into contact with conducting objects or liquids Check the prod...

Page 11: ...from the application area Only start the application when combustible or flammable gases and liquids are no longer present Only connect or disconnect the power plug to or from the power supply outside...

Page 12: ...CKLINE working insert 6 HF connection 7 Locking button for releasing the working insert with outer sheath 8 Rotating wheel for turning the jaws 9 LUER Lock connector for irrigation with sealing cap 10...

Page 13: ...refer to the Technical data chapter for this in formation Instrument access sys tems To maintain endoscopic access to the patient the outer sheath including a working insert can be inserted percutane...

Page 14: ...instructions for use The color may differ on the product The symbol is black white on the packaging label Unsterile Temperature limit Humidity limit Federal USA law restricts this device to sale by o...

Page 15: ...0 mm TBF107_EN_V1 0_10 2021_IFU_CE MDR 15 4 5 Ambient conditions Storage transport conditions Temperature 29 C 50 C 20 F 122 F Relative humidity non condensing 5 85 Operating conditions Ambient temper...

Page 16: ...eries document their nature and extent and contact the manufacturer or supplier immediately 5 2 Inspecting the device 1 Inspect the device for visible contamination Do not use if contaminated 2 Inspec...

Page 17: ...e the outer sheath with closed jaws into the completely opened handle as far as it will go Close the handle completely The outer sheath is now locked in the handle The CLICKLINE instrument is now read...

Page 18: ...d part or defibrillation proof type CF applied part Residues of cleaning agents disinfectants solvents or flammable gases may accumulate on the underside of the patient s body and in body cavities or...

Page 19: ...s Risk of burns Burns to the user Let the HF electrodes cool down after use 1 Open the handle 2 Press and hold the locking button do not turn 3 Open the handle to the maximum in order to pull the oute...

Page 20: ...truction manual are exempt from this rule Please contact your local KARL STORZ subsidiary or authorized dealer see the list of subsidiaries Contaminated devices may not be shipped To prevent contact i...

Page 21: ...e vices 26005 M Unipolar high frequency cable with 5 mm plug length 300 cm for use with AUTOCON II 400 SCB system 111 115 122 125 AUTOCON III 300 400 AUTOCON II 200 AUTOCON II 80 KARL STORZ AUTOCON sy...

Page 22: ...ns for use CLICKLINE Instruments 10 mm TBF107_EN_V1 0_10 2021_IFU_CE MDR 22 9 2 Spare parts Art no Designation 29100 Sealing cap for LUER lock irrigation connector black autoclavable package of 10 790...

Page 23: ...eath 10 mm 36 cm uninsulated metal Handle accessories 30131 33121 33500 33500CM 33120 30133 33121W 33500G 33500M 33120G 33131 33122 33500W 33120T 33131P 33123 33700 33131PSR 33126 33131T 33136P 33131W...

Page 24: ...6 cm HF forceps insert with reinforced bay onet HF scissor insert with reinforced bay onet 33510AF 33510BC 33510BL 33510BLS 33510CB 33510DU 33510FK 33510FS 33510GF 33510H 33510KA 33510MLL 33510PG 3351...

Page 25: ...rz se KARL STORZ Endoscopy Suomi OY Taivaltie 5 01610 Vantaa Finland Phone 358 0 96824774 Fax 358 0 968247755 E mail asiakaspalvelu karlstorz fi KARL STORZ SE Co KG Representative Office algirio St 94...

Page 26: ...07 Building B 18F Beijing IFC No 8 Jianguomenwai Street Chaoyang District 100022 Beijing People s Republic of China Phone 86 10 5638188 Fax 86 10 5638199 E mail info karlstorz com cn KARL STORZ Endosc...

Page 27: ...Subsidiaries Instructions for use CLICKLINE Instruments 10 mm TBF107_EN_V1 0_10 2021_IFU_CE MDR 27...

Page 28: ...RZ SE Co KG Dr Karl Storz Stra e 34 78532 Tuttlingen Postfach 230 78503 Tuttlingen Germany Phone 49 7461 708 0 Fax 49 7461 708 105 E mail info karlstorz com www karlstorz com TBF107 EN V1 0 10 2021 IF...