Reprocessing
Instruction For Use • C-MAC
®
Video Laryngoscope 8403xxx and Connection Cable 8403X • 96076019EN-US •
V4.3
26
CAUTION
For the production and application of solutions, the chemical manufacturer's
specifications regarding concentration, exposure time, and life span must be strictly
adhered to. An excess exposure time or incorrect concentration can lead to damage.
Bear in mind the microbiological range of action of the chemicals used.
CAUTION
Damage to the product!
Only perform reprocessing using chemicals approved by KARL STORZ. A list of approved
chemicals can be requested from [email protected].
As a general rule, all system components 8403 xxx C-MAC
®
video laryngoscopes and C-
MAC
®
connection cable 8403 X are resistant to manual and mechanical cleaning and
disinfection. If the optional protective cap is not used, particular care must be taken to
ensure that the contacts are dry so as not to impair functionality.
ATTENTION
Observe the national regulations and A
0
value.