![Storz C-MAC 8401 Series Instructions For Use Manual Download Page 18](http://html1.mh-extra.com/html/storz/c-mac-8401-series/c-mac-8401-series_instructions-for-use-manual_1382921018.webp)
Installation and commissioning
Instruction For Use • C-MAC
®
video laryngoscope 8401xxx, Electronic Module 8401/8402 X, C-MAC
®
PM 8401
XD • 96076011EN-US • V3.1
18
The lithium ion battery can be used for approx. 1 hour when fully charged.
WARNING
Low battery
If the charge status symbol flashes red, the charge of the C-MAC
®
PM is too low. In this state,
it will only function for another 10 minutes.
Connect the C-MAC
®
PM to the corresponding charge station (8401 XDL) when the charge
status symbol flashes red at the latest.
For the C-MAC
®
PM there is a specially designed bag 8402 YE available, which can be
ordered as an option (see chapter 15.1, Accessories [ 34]). The C-MAC
®
PM is also
included in a set which includes several different C-MAC
®
video laryngoscopes. The article
number is 8400 B "Intubation Set -C22-, Model ULM".
5.2.3 Putting the electronic module into service
Push the electronic module (2) into the receptacle of the video laryngoscope.
The electronic module can be removed from the laryngoscope and inserted into another
laryngoscope during operation.
2
5.2.4 Focus
When using C-MAC
®
video laryngoscope blades, the image is focused automatically. Manual
focusing is not possible/necessary.