6 Operation
6.1 Basic information on operation
WARNING
If the rotating wheel is not secured to the handle, the patient can be injured by
uncontrolled rotations of the shaver attachment.
WARNING
Only operate the handpiece when the applied parts are visible.
WARNING
When using rotating instruments, ensure that the instruments are sufficiently irrigated
and cooled. Inadequate irrigation can result in contaminated or blocked instruments, or
cause jerky movements. This can generate excessive heat and endanger the user or
patient.
WARNING
Risk of burns. A system that is not operated intermittently when working with the motor
or handpiece can increase the instrument, handpiece, and motor temperature.
Checking the system combination before use
Check the irrigation and suction function in vitro before using the system combination. To do
this, connect the DRILLCUT-X II shaver handpiece with suction drill/shaver attachment to the
control unit and activate the suction/irrigation function. (Refer to the instructions for use for the
control unit 96056025D, 96600526D or 960260003D used).
Rinsing the shaver/drill attachments after use
Rinse the suction drill/shaver attachment with the DRILLCUT X II handpiece (with the cutting
window open) immediately after use by activating the suction function, preferably with distilled
water, in order to prevent the attachment from becoming blocked.