For attachment to the medical rail it may be necessary to move the uni-
versal attachment device into another position. For this the two
mounting
screws
must be released. After changing the position, the screws must
be screwed again according to the positioning of the attachment device.
9 Getting started
WARNING
Risk of electric shock!
Before every use, check to make sure that the ASTOPAD DUO310 control unit
and the applied parts are undamaged.
WARNING
Risk of infection!
•
Use aseptic procedures.
•
Route the extension connection cable between the applied part and the con-
trol unit in such a way that it is protected from soiling of blood and body fluids.
•
Prevent the cables from coming into contact with the floor.
•
It is recommended to always place a waterproof and absorbent barrier be-
tween the patient and the ASTOPAD applied part.
WARNING
Risk of injury!
•
When RF surgical instruments or endocardial catheters are used, the patient
must also be properly insulated. This insulation must not be damp. The de-
vice manufacturer’s instructions for use must be observed at all times.
•
With transdermal drug applications (patches), the additional heat can
increase the uptake of the drugs and result in injury to the patient.
•
In the case of arterial occlusion, the applied parts of the ASTOPAD may not
be used distal to this area.
•
Overheating of ischemic extremities can occur with the use of the ASTOPAD
applied parts.
•
When applied parts ASTOPAD COV are used as a top blanket, ensure that
they do not obstruct the patient’s field of vision.
•
Have the device repaired if the self-test is not activated automatically after the
mains switch is switched on and the heating does start immediately.
•
The extension connection cable and the mains cable should not touch the
patient and should not hinder the treating personnel.