2-13
764333-675
STABILIZING LOW
(THERMOCOUPLE LOCATION) = XX.XC
Display shows:
LO SET PT XX.X
PRESS SV TO SET
Using key pad, adjust displayed temperature to match
temperature measured by calibrated device, then press
SAVE VALUES
(SV) touch pad. Display shows:
PLACE TC(S) IN
HI BATH PRESS SV
Place thermocouple(s) and temperature measuring
device in hot water container and press
SAVE
VALUES
touch pad. Display shows:
STABILIZING HIGH
(THERMOCOUPLE LOCATION) = XX.XC
Display shows:
HI SET PT XX.X
PRESS SV TO SET
Using keypad, adjust displayed temperature to match
temperature measured by calibrated device, then press
SAVE VALUES
touch pad. Display shows:
* UPDATING *
* MEMORY *
Display shows:
* (THERMOCOUPLE LOCATION) *
* CALIBRATED *
Printer then prints calibration data. See Figure 2-7, for
example. Press
CANCEL
touch pad to exit calibration.
Display shows: MFG TESTS VALUES CALIB
• Concentration Probe Calibration (CONC)
This routine is used to calibrate Control Board using
established values. Technician is required to access test
points on Connector Board adjacent to Power Supply.
From Service Menu, Select
CALIB,
then press
SAVE
VALUES
touch pad. Display shows:
CALIB RESTORE
VIEW PRINT
Select
CALIB
, then press
SAVE VALUES
touch pad.
Display shows:
PRESS TEMP
CONC MORE
Select
CONC
, then press
SAVE VALUES
touch pad.
Display shows:
WARNING! CALIB
WILL BE CHANGED
At this point, CN9 must be disconnected for the
remainder of this calibration procedure. Press
SAVE
VALUES
touch pad. Display shows:
PLACE JUMPER ON
TP 1 & 2 PRESS SV
Printer prints existing CONC calibration data.
Locate test points on connector board and jumper TP1
and TP2 (low points). Press
SAVE VALUES
touch pad.
Display shows:
STABILIZING LOW
CONC = XX
Summary of Contents for SYSTEM 1E
Page 1: ...MAINTENANCE MANUAL SYSTEM 1E Liquid Chemical Sterilant Processing System 12 09 10 P764333 674 ...
Page 3: ......
Page 71: ...5 19 764333 675 Figure 5 2 Analog Cable Assembly Pin Out Locations ...
Page 74: ...5 22 764333 675 Figure 5 7 DC Power Supply Harness Pin Out Locations PS P1 ...
Page 96: ...7 19 764333 675 Figure 7 17 Aquafine STERIUV5 Control Panel The Control Panel A B 3 2 1 ...
Page 108: ......
Page 110: ...ii 764333 676 Example 1 How To Use IPB Page Typical ...
Page 119: ...10 764333 676 Figure 1 SYSTEM 1E Piping Schematic ...
Page 121: ...12 764333 676 Figure 2 SYSTEM 1E Complete 4 3 2 ...
Page 128: ...19 764333 676 Figure 5 Fluid and Air Components 2 of 2 51 53 48 47 50 49 52 ...
Page 133: ...24 764333 676 Figure 7 Float Block Switch Assembly 1 5 8 7 5 4 10 4 2 6 9 9 3 CK2 CK3 LS3 CK1 ...
Page 137: ...28 764333 676 Figure 9 Drain Block Assembly 3 4 5 6 7 3 1 2 8 CK8 TC3 ...
Page 143: ...34 764333 676 Figure 12 Bulkhead Connector Assembly 1 2 2 2 3 1 1 3 5 4 4 6 ...
Page 155: ...46 764333 676 Figure 18 Harness Cable Assemblies 1 of 2 1 2 3 4 5 6 P10 PT1 P5B P11 P2 P5A ...
Page 166: ......
Page 167: ......
Page 168: ......
Page 169: ......
Page 170: ......