4-47
Processing Instructions
Operator Manual
920011-649
4.13 Responding to a
Fault
Fault messages warn operator if the processor is experiencing an
abnormal condition. Fault conditions can be caused by failure of
utility supplies or by processor components.
Refer to
for a detailed listing of fault
alarms.
NOTE:
Some fault conditions can be solved by the operator.
4.13.1 Faults
When a malfunction or fault occurs before the cycle is successfully
completed:
1. An alarm sounds and display shows:
alternating with:
and printer records:
=========================
HH:MM:SS XM
MM-DD-YY
=========================
= FAULT XX
=
=========================
2. Press
ALARM REPLY
touch pad to acknowledge alarm. This
will record time at which operator acknowledged alarm. Printer
records:
=======================
HH:MM:SS XM
MM-DD-YY
=======================
ALARM ACKNOWLEDGED
=======================
If
FAULT
alarm requires no further action (for example, refer to
), or if the operator
has been able to solve the fault condition using
, display shows:
3. Press
START
touch pad to resume cycle operation. Display
shows:
FAULT XX
CYCLE DELAYED
SEE FAULT CHART
CYCLE DELAYED
START TO RESUME
STOP TO CANCEL
OPERATING
MMM MIN
WARNING
–
INEFFECTIVE
PROCESS HAZARD: Endoscopes
and
accessories
are
not
adequately processed, rinsed
and
air
purged
when
processing cycle is aborted.
REPROCESS LOAD.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...