4-39
Processing Instructions
Operator Manual
920011-649
If
DUPLICATE PRINTOUT
feature is enabled in Miscellaneous
Values menu of Supervisor mode (refer to
, for setting procedure), display shows:
5. Press
SCROLL arrows
(up or down) to choose (
YES
; generates
a duplicate of last cycle printout) or (
NO
; do not generate a
duplicate of last cycle printout).
NOTE:
If MULTIPLE PRINTOUTS feature is enabled, the
duplicated printout(s) is (are) recorded at the end of all printouts.
Header is identified as follows:
========================
DUPLICATE PRINT
# X/X
========================
6. Press
ALARM REPLY
touch pad to silence audible alarm.
Proceed with
.
4.9 Process
Verification and
Unloading Procedure
Reliance EPS Processor
provides the operator with methods of
quality assurance monitoring and allows for immediate parametric,
process and physical verification of the cycle.
IMPORTANT: Good Hospital Practice dictates processing of all
endoscopes be verified after high level disinfection in the Reliance
EPS. Should any processing cycle fail verification, endoscope(s)
must be reprocessed in another cycle.
•
Parametric controls
are performed by the Reliance EPS
Processor to ensure each cycle meets specifications and cycle
data is documented on the printed tape (see Figures 4-24 and 4-
25).
•
Process verification
is achieved through use of a STERIS
chemical indicator and fresh dry chemistry container detection
meter.
•
Physical verification
of the process is performed by the operator.
The operator must ensure all process parameters were acceptable
as explained in
S
ECTION
4.9.1, VERIFICATION 1
–
P
ARAMETRIC
C
ONTROLS
, through
S
ECTION
4.9.3, VERIFICATION 3
–
P
HYSICAL
.
IMPORTANT: Processing verification must be performed within
five minutes after opening the chamber door.
, to
ensure all verification steps are completed as described in
S
ECTION
4.9.1, VERIFICATION 1
–
P
ARAMETRIC
C
ONTROLS
, through
S
ECTION
4.9.3, VERIFICATION 3
–
P
HYSICAL
.
DUPLICATE PRINT
YES
WARNING – INEFFECTIVE
PROCESS HAZARD:
•
Correct function of the
system is only ensured
after any diagnosed errors
have
been
found,
corrected and a successful
processing cycle has been
completed.
•
When uncertain about any
endoscope
processing
cycle, reprocess load.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...