7-2
920011-649
Operator Manual
Troubleshooting
7.1
Troubleshooting
Char
t
PROBLEM
POSSIBLE CAUSE AND CORRECTION
1. Interior light does not turn
ON
.
1. Unit is in
OFF
position; press top portion of
POWER
(ON/OFF)
toggle switch to turn
ON
processor.
2. Interior light bulb is burned out. Replace light bulb. If faulty, call
STERIS.*
2. Cycle does not start when
START
touch pad is pressed.
1. Door is not closed; push door firmly to close and press
START
touch pad.
2. If condition reoccurs, call STERIS.*
3. Little or no water through rotary
spray arms or manifolds.
1. Processor rotary spray arms are clogged or obstructed; clean
spray arms (refer to
S
ECTION
).
2. Endoscope processing support is not correctly positioned in
chamber. Open door, push support firmly into back of chamber.
3. Chamber debris screen may be clogged; clean if necessary.
4.
ENDOSCOPES AND ACCESSORIES MUST BE REPROCESSED
THROUGH A COMPLETE RELIANCE EPS CYCLE
(refer to
).
5. If condition reoccurs, call STERIS.*
4. Rotary spray arms are not
turning.
1. Bushing not properly installed or missing on rotary spray arm hub
(refer to
S
ECTION
).
2. Rotary spray arms are clogged; clean spray arms (refer to
S
ECTION
).
3.
ENDOSCOPES AND ACCESSORIES MUST BE REPROCESSED
THROUGH A COMPLETE RELIANCE EPS CYCLE
(refer to
).
4. If condition reoccurs, call STERIS.*
5. Water leaks from processor.
Call STERIS.*
6. Chamber
does
not
drain
completely.
1. Chamber debris screen may be clogged; clean if necessary.
2.
ENDOSCOPES AND ACCESSORIES MUST BE REPROCESSED
THROUGH A COMPLETE RELIANCE EPS CYCLE
(refer to
).
3. If condition reoccurs, call STERIS.*
7. Vapor, mist or unusual odor
emanating from chamber.
Damaged door gasket and/or damaged door interlocks, call STERIS.*
8. Heat loss from chamber. Hot air
escaping from door during air
purge phase.
Damaged door gasket and/or door interlocks, call STERIS.*
*Service charges may be incurred. Consult your warranty for details.
Summary of Contents for P920011-649
Page 1: ...OPERATOR MANUAL Reliance EPS Endoscope Processing System 2022 03 10 P920011 649 ...
Page 2: ......
Page 13: ...xi 920011 649 Operator Manual Table of Contents ...
Page 17: ...xiv 920011 649 Operator Manual List of Tables Healthcare Capital Equipment ...
Page 25: ...1 8 920011 649 Operator Manual Safety Precautions Healthcare Capital Equipment ...
Page 45: ...3 16 920011 649 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 167: ......
Page 168: ...STERIS Corporation 5960 Heisley Road Mentor OH 44060 1834 USA 440 354 2600 800 548 4873 ...