6-2
10043124
Operator Manual
Maintenance
6.3 Inspect Wall
Control
1.
Check wall control operation by cycling it OFF and ON several
times.
2.
Cycle through intensity cycle for each lighthead controlled by
the system.
a.
Verify proper operation of all membrane switches (buttons).
b. Verify function of all video camera commands if the system is
equipped with an optional video command menus.
6.4 Optional Service
Disconnect Switch
An optional switch (see
) is available to interrupt the ac
voltage to the lighting system (and also dc battery backup voltage, if
the latter is available).
•
This switch is referred to as a
service disconnect switch
and its
purpose is to provide a method of temporarily removing power
from the STERIS Lighting System without having to access the
ON/OFF switch inside the canopy. It is important that you refer to
I
NSTALLATION
I
NSTRUCTIONS
, P129388-228,
included in the
Optional Service Disconnect Switch Kit (P146670-049) for wiring
details.
•
The switch is mounted remotely, in a location chosen by the
Customer.
Important:
The service disconnect switch should not be used to
routinely turn the system on and off.
Figure 6-1. Optional Service Disconnect Switch
Summary of Contents for HarmonyAIR G Series
Page 1: ...OPERATOR MANUAL HarmonyAIR Surgical Lighting System G Series Rev G 10043124 ...
Page 2: ......
Page 8: ...vi 10043124 Operator Manual Table of Contents Healthcare Capital Equipment ...
Page 10: ...ix Table of Contents Operator Manual 10043124 Healthcare Capital Equipment ...
Page 24: ...2 6 10043124 Operator Manual Installation Verification Healthcare Capital Equipment ...
Page 42: ...3 18 10043124 Operator Manual Operating Instructions Healthcare Capital Equipment ...
Page 46: ...4 4 10043124 Operator Manual Cleaning The Equipment Healthcare Capital Equipment ...
Page 54: ...7 2 10043124 Operator Manual Replacement Parts Healthcare Capital Equipment ...
Page 56: ...8 2 10043124 Operator Manual Waste Disposal Guidelines Healthcare Capital Equipment ...