7-1
Cycle and Control Value Programming
Operator Manual
P129394-020
AMSCO
®
400 Series Steam Sterilizers are shipped with factory-set
cycles, cycle values and control values programmed into the control
(see
and
page ii and v). Preset values can
be changed to tailor the sterilizer to the operating environment in
which it has been placed. To change these values, refer to
NOTE: If a control board failure should occur, operator-set cycles and
values may be lost.
Important:
Applicable cycles have been validated to satisfy the
requirements outlined in
. If cycle parameters
(sterilize time, dry time, temperature) other than those listed are
required, it is the responsibility of the healthcare facility to consult
and follow the device manufacturer’s written instructions.
NOTE: Although AMSCO 400 Series Steam Sterilizers are factory-
set to operate using Fahrenheit temperature units, the sterilizer can
be reprogrammed to display and print using Celsius temperature
units.
7.1 Change Values
Procedure
The change values procedure described in the following pages can
be used to adjust cycle values within a limited range and to make
selections affecting operation of the unit. Refer to
Tables
and
for tested cycle values.
NOTE: Default cycle values are shown on
Tables
and
The Change Cycle Values procedure can be used to adjust the
following values:
•
Name: The user can select between 3 to 4 factory-programmed
names (depending on sterilizer configuration), and can create
up to two user-defined cycle names.
•
Sterilization Exposure Temperature: Exposure temperatures can
be adjusted, but within limits. Refer to
for details.
•
Sterilization Exposure Time: (
Important:
This value is limited to a
minimum value, refer to
). Refer to
for range limits.
•
Dry Time: Refer to
for range limits.
NOTE: Change Values can also be used to make adjustments to the
Change Machine Setup values shown on
The Change Values touch screens are accessed by pressing the
Change Cycle Values touchpad (see Screen #2). All change values
options can be secured (or locked out) using a supervisor’s access
code. It is recommended that supervisor’s Access Code be used to
prevent unauthorized personnel from changing cycle and system
setup parameters.
NOTE: Although AMSCO 400 Series Sterilizers are factory-set to
operate using Fahrenheit temperature units, the sterilizer can be
reprogrammed to display and print using Celsius temperature units.
IMPORTANT
Applicable cycles have been validated
to satisfy the requirements outlined in
and
. If cycle
parameters (sterilize time, dry time,
temperature) other than those listed
are required, it is the responsibility of
the healthcare facility to consult and
follow the device manufacturer’s
written instructions.
CYCLE AND CONTROL VALUE PROGRAMMING
7
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...