![Steris AMSCO 400 Series Operator'S Manual Download Page 34](http://html.mh-extra.com/html/steris/amsco-400-series/amsco-400-series_operators-manual_1374315034.webp)
4-4
P129394-020
Operator Manual
Techniques of Sterilization
4.7 Techniques of
Sterilization for
Liquid Cycle
Important:
The liquid cycle is for non-patient contact use only.
for recommended Liquid cycle parameters. The
recommended times indicated in
assume the use of vented
bottles. The “minimum recommended sterilize time” includes the
time required to bring the solution up to the sterilize temperature plus
the time required to achieve sterilization.
WARNING – EXPLOSION HAZARD: This sterilizer is not designed to process flammable
compounds.
WARNING – PERSONAL INJURY HAZARD: Avoid personal injury from bursting bottles. Liquid
sterilization cycle must only be used for liquids in borosilicate (Pyrex) flasks with vented closures.
WARNING: The liquid cycle is for non-patient contact use only.
WARNING – BURN HAZARD: When sterilizing liquids, you must observe the following procedures:
•
Use Liquid cycle only.
•
Use only vented closures.
•
Use only Type I borosilicate glass bottles.
•
Processed bottles are hot following liquid cycle and may boil over. Always wear appropriate PPE
including thermal gloves and goggles when removing processed bottles from chamber.
Do not allow
hot bottles to be jolted.
• Steam may be released from the chamber when door is opened. Step back from the sterilizer
each time the door is opened to minimize contact with steam vapor.
CAUTION – POSSIBLE EQUIPMENT DAMAGE: Sterilization of chloride-containing solutions (e.g.,
saline) can cause chamber corrosion and is not recommended by the manufacturer. If, however,
chloride-containing solutions must be processed, clean the chamber after each use.
!
!
!
!
!
Table 4-1.
Liquid Cycle Processing Guidelines
Sterilizer Size
Number of
Containers
for Full
Load
Volume of
Liquid in
One
Container
Minimum
Recommended
Sterilize Time
at 250°F
(121°C) in
minutes
16 x 16 x 26"
(406 x 406 x 660 mm)
15
1000 mL
45
20 x 20 x 38”
(508 x 508 x 965 mm)
with Loading Car
32
20 x 20 x 38”
(508 x 508 x 965 mm)
with Rack and Shelves
28
Summary of Contents for AMSCO 400 Series
Page 2: ......
Page 10: ...viii P129394 020 Operator Manual Introduction Healthcare Capital Equipment ...
Page 18: ...xvi P129394 020 Operator Manual List of Figures and Tables Healthcare Capital Equipment ...
Page 28: ...2 6 P129394 020 Operator Manual Symbol and Labels Healthcare Capital Equipment ...
Page 60: ...5 24 P129394 020 Operator Manual Component Identification Healthcare Capital Equipment ...
Page 69: ...6 9 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 73: ...6 13 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 77: ...6 17 Sterilizer Operation Operator Manual P129394 020 Healthcare Capital Equipment ...
Page 142: ...8 12 P129394 020 Operator Manual Routine Maintenance Healthcare Capital Equipment ...
Page 174: ...9 32 P129394 020 Operator Manual Troubleshooting Healthcare Capital Equipment ...