5-7
764332-898
5.2 ELECTRICAL TROUBLESHOOTING
See Troubleshooting Guide (
) for a sequence
of steps to use to identify and correct most electrical
problems encountered. Also refer to the electrical
schematics behind
tab when
using the Troubleshooting Guide.
, when the
cause of the problem appears to be hydraulic.
, Battery Diagnostics, to determine
proper battery voltages (battery-operated tables only).
Electrical Troubleshooting Guidelines:
• Before replacing any PC boards or other components
suspected of being defective, ensure all cable plugs
are fully inserted into their respective jacks.
• Ensure all cable pins are installed and locked into
place on the plug(s).
• When making any dc voltage measurements,
reference the dc ground, not the chassis ground.
(These two points are isolated from each other.)
The pump motor and/or the hydraulic solenoids
can be checked by applying 24 Vdc only directly to
the component.
• To completely disconnect power to the table, the line
cord as well as the batteries must be disconnected.
• When resoldering or replacing any of the table or
floor-lock microswitches, isolate the control by
disconnecting its associated plug.
Figure 5-1. Amsco 3085 SP Surgical Table Block Diagram
26. Table articulates with NO
input from hand control or
override command.
Electrical system
malfunction
Faulty override switch board from liquid
exposure - replace override switch board
and verify RTV is applied around bellows
for fluid intrusion protection.
Table 5-1. Troubleshooting Guide (Cont’d)
Situation
Cause
Service Instructions
Table 5-2. Battery Diagnostics
Battery
Low Battery LED Status
ON
Flashing
Motor
22.7 Vdc
21.6 Vdc
Control
23.6 Vdc
22.7 Vdc
CAUTION – POSSIBLE EQUIPMENT
DAMAGE: Isolate the coil and/or pump
motor from the power control by
disconnecting the wires on the motor
terminal board, or slip the wires off the
solenoid coil that is being worked on.
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......