6-2
764332-898
Figure 6-1. Hydraulic Pump Relief Valve
6.4 TABLE OBSERVATION
1. Check entire table for proper fit and finish of all
exposed parts. Check for any burrs, sharp edges or
corners to which users could be exposed. Correct as
necessary.
2. Check base/column shrouds for damage and
proper movement/engagement.
3. Using a Clark socket, ensure socket slides freely
over each side rail and that rivet stops and gravity
locks prevent socket from being removed. Ensure
each gravity lock swings freely.
4. Flip foot pump pedal up and down several times to
ensure it pivots smoothly and stays in position.
5. Articulate headrest throughout its range (+90° to
-90°). Ensure ratchet action is normal and release
handle operation is smooth throughout range.
6. Raise kidney bridge to maximum height (top
surface of bridge should be at least 3-7/8" [98 mm]
from top surface of leveled back section). Ensure
ratchet action is smooth and normal throughout.
Flip handle into and out of its stowed position
several times, ensure detent retains handle in
stowed position.
6.5 TABLE MOBILITY
NOTE: The following tests are to be done with the floor
locks disengaged, with the tabletop horizontal and at
maximum elevation, and on a smooth level surface.
1. Move table forward, backward, left and right: 3"
(914 mm) minimum in each direction. Ensure
casters operate smoothly and without unusual
noise.
2. Move table longitudinally, with headrest leading,
until casters are aligned for movement in that
direction. Attach a spring scale to end of leg section
on table centerline. Slowly push table straight
ahead. Gauge reading should be less than 35 lb
(16 kg).
3. With table in same alignment, slowly push at right
angles to table centerline, at outboard end of left
side rail on table headrest. Gauge reading should
be less than 20 lbs (9 kg).
6.6 FLOOR LOCK ASSEMBLY
1. With table on a level floor, engage floor locks.
2. Check distance between each caster and floor: it
should be 1/4 ±1/32" (6 ±0.8 mm). If adjustment is
necessary, refer to
.
6.7 TABLE ARTICULATION
1. Articulate each table function through entire range
and ensure there are no unusual noises, speed
variations or other performance problems. Table
speed and range of movement should fall within
parameters shown in
2. Actuate RAISE function for at least 30 seconds. This
causes pump to develop maximum pressure and
relief valve to be fully employed. There should be
no chattering, squealing or other unusual noises.
(for electric table) and
(for battery-powered table) and articulate table
through various functions using conditions
described.
NOTE: If a battery-operated table, the base shroud
must be removed for the last three conditions - see
Step 4
.
4. Remove table column and base shrouds as
5. Thoroughly examine table assembly for evidence of
any hydraulic leaks, mechanical wear, electrical
wire abrasion or any other degradation. Articulate
each table function again through entire range and
ensure there are no unusual noises, speed variations
or other performance problems.
Relief Valve
Adj
us
ti
n
g
S
c
r
ew
Relief Valve
Adjusting Screw
Summary of Contents for Amsco 3085 SP
Page 1: ...MAINTENANCE MANUAL Amsco 3085 SP Surgical Table After S N B420702 025 10 07 10 P764332 897 ...
Page 3: ...OPERATOR MANUAL Amsco 3085 SP Surgical Table 08 10 11 P150830 353 ...
Page 56: ......
Page 132: ......
Page 155: ...18 764332 899 Figure 5 AC Plate Assembly 1 7 3 4 5 6 2 9 10 P17 ...
Page 161: ...24 764332 899 Figure 8 Column Assembly 1 2 3 4 5 6 7 8 9 ...
Page 176: ...40 764332 899 Figure 15 Leg Section Assemblies 1 7 2 4 3 5 6 ...
Page 204: ......
Page 205: ......
Page 206: ......
Page 207: ......
Page 208: ......
Page 209: ......
Page 210: ......
Page 211: ......
Page 212: ......
Page 213: ......
Page 214: ......
Page 215: ......
Page 216: ......
Page 217: ......
Page 218: ......
Page 219: ......
Page 220: ......
Page 221: ......
Page 222: ......
Page 223: ......
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......