4-12
150832-639
Operator Manual
Operating Instructions
4.8 X-Ray Top
Installation
The STERIS 4085 General Surgical Table is designed to enable
installation of an X-ray top to allow use of X-ray cassette film. Attach
the four-part X-ray top as follows (see Figure 4-9):
1.
Carefully remove X-ray top sections from packing container.
2.
Remove pads from tabletop (see
).
3.
Position X-ray top section over corresponding tabletop section,
aligning posts to mounting holes.
4.
Press down on X-ray top section until a "click" is heard.
5.
Ensure X-ray top sections are securely mounted by gently
pulling up on section.
6.
Reinstall tabletop pads.
7.
Remove and store X-ray top by reversing above steps.
Figure 4-9. X-Ray Top Installation (Typical)
Ensure caps cover
sockets when not in use.
SERVE
C-MAX
10
2
1
CLICK
LLING -RA
TOP
p
ds and stow it with care.