Figure 6-1. Table Sensor Calibration
8. Display shows
ADJUST MODE.
NOTE: During table sensor calibration, note the
following:
1) The
Reverse Patient
LED must be OFF.
2) Do not press
LEVEL
button during procedure
or control exits Calibration Mode
without
storing the values.
3) Pressing
FLOOR LOCK
button during
procedure orders the control to skip displayed
articulation. This is NOT recommended during
a final calibration.
9. If
ADJUST MODE
does not display or
Reverse
Orientation
LED is ON, turn Hand Control OFF
(press
STOP
) then ON and repeat
Step 7
until
correct results appear.
10. Press
Reverse Orientation
button, display shows
LEVEL ADJ
.
11. Using a level, position tabletop (starting with seat
section) so all sections are level in both horizontal
directions.
12. Slide tabletop until it is approximately 3/16"
(4.8 mm) past foot end of tilt frame (see
Figure 6-1). This is known as "Zero Slide" position.
13. Press
Reverse Orientation
button to save setting.
14. With display showing
TILT LEFT MAX
, tilt
tabletop to maximum left tilt position and press
Reverse Orientation
to save setting.
15. With display showing
TILT RIGHT MAX
, tilt
tabletop to maximum right tilt position and press
Reverse Orientation
to save setting.
16. Display shows
TREND MAX
. Use
TILT LEFT
button to return table to center. Do not use
LEVEL
button.
17. Put table into maximum Trendelenburg then
press
Reverse Orientation
button to save setting.
18. Display shows
REV TREND MAX
. Put table into
maximum Reverse Trendelenburg then press
Reverse Orientation
button to save setting.
19. Display shows
LEG DOWN MAX
. Use
TREND
button to return table to center. Do not use
LEVEL
button.
20. Lower leg section to maximum position and press
Reverse Orientation
button to save setting.
21. Display shows
LEG UP MAX
. Raise leg section to
maximum position and press
Reverse Orientation
button to save setting.
Tilt frame
Extend table
top 3/16”
beyond the
foot end of the
tilt frame.
Tabletop
Tabletop
Extend Tabletop
3/16" (4.8 mm)
Beyond Foot
End of
Tilt Frame
Tilt Frame
Summary of Contents for 4085
Page 1: ...MAINTENANCE MANUAL STERIS 4085 General Surgical Table 07 01 10 P764333 765 ...
Page 3: ...OPERATOR MANUAL STERIS 4085 General Surgical Table 04 15 09 P150832 639 ...
Page 131: ......
Page 172: ......
Page 173: ......
Page 174: ......
Page 175: ......
Page 176: ......
Page 177: ......