8 Care and maintenance
©
F. Stephan GmbH
GA-602-0106V1.0-PAE-GB
29
8
Care and maintenance
8.1
Disinfection and sterilisation
As with operation of the device, it must only be cleaned and disinfected
by instructed members of staff.
Proceed with routine cleaning at regular intervals according to the local
hospital routine.
All disposable parts must be disposed of in an environment-friendly
manner according to the local hospital routine.
All instrument and surface disinfectants based on alcohol and aldehyde
may be used.
Do not use deoxygenating or dechlorinating disinfectants or disinfectants
containing phenol or its derivatives.
Amino derivatives can damage soft plastics (tubes).
All used disinfectants must be certified in accordance with relevant
environmental legislation.
Always comply with the regulations for handling cleaning agents and
disinfectants issued by the Professional Association.
When using cleaning agents and disinfectants, pay attention to correct
concentrations and reaction times as otherwise damage to the materials
cannot be ruled out.
When using effective substances other than those stated, please contact
the manufacturer of the disinfectant to obtain compatibility confirmation.