Error Codes 

Error Code 5:

Vial cannot be seated properly (e.g. 

too wide, too short, etc.).

Solution: 

Gently remove the vial. Instrument 

will return to “ready” state and 

display a single light at the top. 

(1) Check for blockage in the 

vial grasping mechanism. Invert 

instrument to gently shake out 

Error Code 7:

“Abandoned” vial left in instrument 

more than 15 seconds after thawing 

process is completed.

Error Code 9:

Vial fails to rise after thaw process 

is finished.

Solution: 

Grasp the vial cap only with fingers 

or forceps and rock vial gently 

to dislodge and remove from 

opening, taking care not to open 

the cap. Instrument will return to 

obstructions, if any. (2) Ensure that 

the thermal pads lining the internal 

heating blocks are intact and 

undamaged. (3) Ensure vial is within 

specifications. (4) Try to thaw frozen 

ThawSTAR

®

 CFT2 Confirmation Vial. 

If thermal pads are not intact and/

or damaged, or if situation is still not 

resolved, contact BioLife technical 

support.

Solution: 

Gently remove vial. Instrument will 

emit a single tone. Instrument will 

return to “ready” state and display a 

single light at the top. 

“ready” state and display a single light 

at the top. Ensure that the thermal 

pads lining the internal heating 

blocks are intact and undamaged. 

Note: If thermal pads are not intact 

and/or damaged, contact BioLife 

technical support. Avoid sticking 

forceps or other foreign objects into 

the opening.

Error Code 2: 

Failure to warm up properly; or oth-

er mechanical fault in the heating 

system.

Error Code 3: 

Vial too warm and/or already 

thawed.

Solution: 

Check power source connection at 

back of instrument, cable connection 

to power adapter and the plug con-

nection to the electrical outlet. Turn 

the instrument off and back on. If the 

instrument displays the same error 

code 2, contact BioLife technical 

support.

Solution: 

Gently remove vial that is too warm. 

Instrument will return to “ready” state 

and display a single light at the top. 