5.  You should be aware of the cleaning procedures  to be used after the use of oil. Wipe oil from  the oil 

immersion  objective,  and  from  any  other  parts  in  need  of  cleaning.  Clean  lenses immediately with 

xylene  after  use.  Leaving  oil  on  the  lens  may  lead  oil  to  seep  behind  and harden on the inside  of 

the lens. 

  To  obtain  a  bright  and  clear  image,  it  is  import  to  adjust  the  illumination system.  The condenser 

position and the iris diaphragm  size should match with objectives  of different magnifications. Turn  the 

condenser  focus  knob  to  shift  the  condenser.  It  needs  to  raise  the condenser  when using the high 

magnification  objective,  and to decline when using the low magnification  one. 

  The  aperture  diaphragm  is  used  to  adjust  the  numerical  aperture  instead  of  brightness. Generally, 

setting  the size of the aperture  diaphragm  at 70%

80%  of the numerical  aperture, you can obtain a 

clear image  with enough  contrast. 

  For observing  a batch  of specimen  of the  same  type,  use  the  limit  bolt  to locate  general position  of 

the  stage  to  simplify  operation.  In  this  way,  the  specimen  can  come  into  focus rapidly  just  by 

turning  the coarse  focusing  knob. 

Note: 

specimens  of  the  same  type  means  they  are  same  in  the  thickness  of  slide,  cover slip and 

specimen  slice. 

 

7. Maintenance

 

1.  Carefully open the box, avoid the accessories,  like lens, dropping  to ground and being damaged. 

2.  All  the  lenses  have  been  well  checked  and  adjusted.  It is  forbidden  to disassemble  them by user 

self. 

3.  The  nosepiece  and  coarse/fine  focus  unit  have  a  compact  and  precise  frame,  so  please don

’t 

disassemble  them as possibly  as you can. 

4.  Keep  the instrument  clean,  wipe  dust  regularly,  apply  a little  non-corrosive  grease  in the sliding 

parts and be attention  to avoid contaminating  the optical elements  especially. 

5.  Place the  instrument in  a  cool,  dry  position. When not  using the  microscope,  keep  it covered  with 

a dust helmet. Make sure the lamp socket is cool before covering  the microscope. 

 

8. Troubleshooting 

Under  certain conditions,  performance  of this unit may be adversely  affected  by factors other than 

defects.  If a problem  occurs,  please  review  the following  list and take remedial action as needed.  If you 

cannot solve the problem after checking  the entire list, please  contact your local franchiser  for assistance. 

 

Summary of Contents for NYMCS-212

Page 1: ...lar microscope NYMCS 212 Toensure the safety obtain optimum performance and to familiarize the use of this microscope it is recommended strongly that you study this manual thoroughly before using the...

Page 2: ...NYMCS 212 is widely used in biological pathological bacteriological fields for purposes such as observation research academic demonstration clinic medical inspection It features more comfortable and...

Page 3: ...198 3 Eyepiece Category Magnification Focal Length f mm Field of view mm Plan Eyepiece 10X 24 95 18 4 Total Magnification Eyepiece 10X 10X 10X 10X Objective 4X 10X 40X 100X Total 40X 100X 400X 1000X 5...

Page 4: ...nosepiece has a precise frame which can locate objectives accurately and stably The digital binocular head can turn 360 degree freely with comforts and conveniences b Curved arm and coarse and fine f...

Page 5: ...the nosepiece clockwise and mount other objectives in the sequence of low to high magnification E Condenser The condenser had been installed in the microscope before leaving factory F Collector The c...

Page 6: ...operation In this way the specimen can come into focus rapidly just by turning the coarse focusing knob Note specimens of the same type means they are same in the thickness of slide cover slip and spe...

Page 7: ...specimen Clean it up The position of the condenser is too low Raise it to proper position The image is defocused low resolution contrast There is no coverslip on the specimen Add coverslip The coversl...

Page 8: ...cked position 2 Mechanical part PROBLEM CAUSE SOLUTION The image can focus when using high magnification objective The specimen is placed inversely The coverslip is too thick Turn inversely Use the st...

Page 9: ...e digital Binocular head with PC by the USB cable the system will find the new device and then run the driver under the CD please follow the clue to finish the installation Attention please link the c...

Page 10: ...ent USB2 0 port requires reinstalling or updating the driver again to make the camera work rightly If not it will influence the performance of the camera and cause the live image displays drag for the...

Page 11: ...esigned for digital microscope It allows you to view capture edit record zoom measure and process microscope images Specific features included are as follows 1 Active image preview freezing preview fi...

Page 12: ...n image zoom in zoom out 1 1 Attention When launching the software there will be several seconds displays as black preview please don t worry it is normal because the camera needs time to do some pret...

Reviews: