23
Press
to enter Auto-Zero option; and press
to change Auto-Zero option.
Auto-Zero can be set to 0 (disabled), 2 (2%F.S.), 4 (4%F.S.), 10 (10%F.S.), 20 (20%F.S.),
50 (50%F.S.).
Press
to enter Manual-Zero option; and press
to change Manual-Zero
option
Manual-Zero can also be set to 0 (disabled), 2 (2%F.S.), 4 (4%F.S.), 10 (10%F.S.), 20
(20%F.S.), 50 (50%F.S.).
Press
to enter Backlight option; and press
to change Backlight option.
Backlight can be set to off (disabled), 5 (5sec), 15 (15sec), 30 (30sec), 60 (60sec), on
(never).
When Backlight is set to 5, 15, 30, or 60, scale turns off backlight in corresponding
seconds after it is stable.
When Backlight is set to off, backlight is disabled, and never turns on
When Backlight is set to on, scale never turns off backlight, no matter it is stable or
not
Press
to exit the Settings Mode.
NOTE: Settings (SETUP) for the scales SBS-KW-300/100-0, SBS-KW-300/100 and SBS-
KW-1TC are similar to the SBS-KW-300AW, SBS-KW-300AB and SBS-KW-300AR sca-
les settings (SETUP), however scales SBS-KW-300/100-0, SBS-KW-300/100 and SBS-
KW-1TC have some additional service settings („dPS”, „oLd” and „dc”) which should
not be changed – displayed values should be confirmed by ZERO|TARE button without
any changes.
24
Item
Operations
Display
Notes
Press
simultaneously,
and keep pressing for 1 second.
SETUP
Press
simultaneously, and keep pressing
for 1 second.
SCALE
Full scale
Press the
button
300
Value 300
should be set
Resolution
Press
to go to resolution
parameter. Press
to change
Resolution.
E 1
Resolution
can be set to
1/2/5
Decimal
Point
Press
to go to decimal
point parameter. Press
to change decimal point.
Pt 0
0: XXXXX
1: XXXX,X
2: XXX,XX
3: XX,XXX
4. CALIBRATION
applicable to scale models SBS-KW-300AW, SBS-KW-300AB and
SBS-KW-300AR
Item
Operations
Display
Notes
Zero Calibration
Press
button
LOAD 0
Load1 Calibration
Press
button
HALF
hang a hook
150 kg
Load2
Calibration
Press
button
FULL
hang a hook
300 kg
Press
button
End
End of calibra-
tion
EN
EN