14
4-4 Care and Maintenance
a. No calibration is needed. Even components exceed the life limitation, there is no
aging problem.
b. Clean the Dental Handpiece
Putting the Dental Handpiece into “PBF” Sonacide Disinfectant (Glutaraldehyde 2%) for
10 minutes under 25
℃
room temperature and cleaning by distilled water. After then,
dipping the Dental Handpiece into 75% alcohol for 5 seconds. Then, wait the Dental
Handpiece for dry.
Step 1
Step 2
Step 3
Step 4
c. Do NOT immerse the device in any other liquid or subject to steam sterilization.
d . Always switch off the device when not use in for prolonged periods of time.
e. The device is not user-serviceable and should never be disassembled.
f. To order any replacement parts, contact the nearest dealer.
g. Any serious incident that has occurred in relation to the device should be reported to the
manufacturer and the competent authority of the Member State in which the user and/or
patient is established