background image

Handbook for the SXVR-H16

Issue 1 June 2009

6

As can be seen above, there is a CCD temperature monitoring window at the right
hand side of the panel. At switch-on, this will default to full power cooling with an
end point of -40C and, needless to say, this is rather extreme. I recommend changing
the set point to about -10C for normal use, but you can go much colder if you are
imaging during the winter months. Under indoor conditions, the low airflow will limit
the cooling capability, and you should use a set point of no lower than -5C for stable
cooling. You can determine the optimum settings for your camera and ambient
conditions when you have some experience of using the system, but do not try to
operate at extreme cooling when the air temperature is high.

Recording your first image:

We now have the camera and computer set up to take pictures, but an optical system
is needed to project an image onto the CCD surface. You could use your telescope,
but this introduces additional complications, which are best avoided at this early
stage. There are two simple options, one of which is available to everyone:

1) Attach a standard ‘M42’ SLR camera lens to the SXVR-H16, using the 27mm

spacer/adaptor to achieve approximately the correct focal distance.

2) Create a ‘Pin hole’ lens by sticking a sheet of aluminium baking foil over the end

of the lens adaptor and pricking its centre with a small pin.

Summary of Contents for SXVR-H16

Page 1: ...a 15mm x 15mm active area The use of high performance microlenses on the CCD surface gives the greatest possible throughput of light to the pixels and the resulting QE is very good over the entire vi...

Page 2: ...ics card capable of displaying an image in a minimum of 1600 x 1200 pixels and 32 bit colour A medium specification Pentium with between 1GHz and 3GHz processor speed is ideal Please note that the SXV...

Page 3: ...ll be able to access it You can confirm that the installation is complete by checking the status of the USB devices in the Windows Device Manager see above Start up the Windows Control Panel and selec...

Page 4: ...nu and image panel appear If this is the first time that it has been run you will receive a warning about the lack of an ini file just click on OK and then open Set program defaults from the File menu...

Page 5: ...click on the camera icon at the top of the screen If the USB connection is OK a message box will inform you of the Handle number for the SXVIO interface and various other version details etc Click OK...

Page 6: ...your camera and ambient conditions when you have some experience of using the system but do not try to operate at extreme cooling when the air temperature is high Recording your first image We now ha...

Page 7: ...or pinhole towards a well lit and clearly defined object some distance away Now click on the camera icon in the toolbar of the SXV H16 software and the camera control panel will appear see above Sele...

Page 8: ...eb site to see if there are any updates or information about your camera software that might help The message board might prove useful to ask for help with getting your camera operating properly Our g...

Page 9: ...opes As a general guide most CCD astronomers try to maintain an image scale of about 2 arc seconds per pixel for deep sky images This matches the telescope resolution to the CCD resolution and avoids...

Page 10: ...an indicator Once you are happy with the focus quality achieved you might like to trim the settings of your par focal or flip mirror eyepiece to match the current camera position Although you can rea...

Page 11: ...ime used during an imaging session As variations in ambient temperature can affect the dark signal it is best to take the dark frames within a few minutes of capturing your images For the same reason...

Page 12: ...Handbook for the SXVR H16 Issue 1 June 2009 12 A 5 minute exposure many warm pixels are visible see the magnified view below...

Page 13: ...he readout noise will be much reduced by the square root of the number of averaged darks and so this is one way to improve the situation but it needs a lot of imaging time to be devoted to gathering t...

Page 14: ...rm pixels are modified the rest of the pixels are unaffected Yet another option is to sum several images that have a small degree of dither in their star alignment If the images are realigned on the s...

Page 15: ...the use of a telescope for deep sky imaging do not forget that you have the option of using an ordinary camera lens for impressive wide field shots A good quality 200mm F3 5 lens with an infrared blo...

Page 16: ...ill not stop until you press a computer key The images will be saved in FITs format with sequential names such as Img23 Img24 and will be found in the Autosave directory or a sub directory of Autosave...

Page 17: ...otes on the camera window and vignetting effects in the optical system of the telescope Dust motes act as inverse pinholes and cast out of focus images of the telescope aperture onto the CCD chip wher...

Page 18: ...y must first have a dark frame subtracted Although this may appear to be unimportant with such brightly lit and short exposures there is the bias offset of the camera in each image and this can produc...

Page 19: ...mbly mounted in front of the SXV camera or with a separate guide telescope rigidly mounted alongside your imaging telescope I personally use it with an 80mm aperture F5 inexpensive refractor as a guid...

Page 20: ...n a 80mm refractor guide scope in the author s garden To use the autoguider please proceed as follows 1 Having started the SXVR H16 software open the autoguider control panel by clicking on the autogu...

Page 21: ...cope and mount You have the option of manually selecting values or asking the software to attempt to determine what they should be This is done by pressing the Train button and waiting for the softwar...

Page 22: ...ation Removing Dust 1 Dust can be deposited on either the optical window not a big problem to cure or on the CCD faceplate difficult to eliminate entirely When small particles collect on the window th...

Page 23: ...isturbed renew it with fresh compound before reassembling 7 Replace all the camera parts in reverse order and the job is done Dealing with condensation The SXVR H16 is designed to avoid condensation b...

Page 24: ...ative to the CCD surface To make an adjustment slacken the appropriate set screw and then turn the adjacent cap head screw in the required direction Complete the adjustment by re tightening the set sc...

Page 25: ...ly be used for normal purposes described in the standard operating instructions and within the relevant safety standards of the country where the equipment is used 2 Repairs under guarantee will be fr...