C A U T I O N
N O T I C E
· During vibration, the motor and the motor cord may affect computer and LAN cable. Noise could be heard during
operation near a radio receiver.
· Turn off the Power Switch after each use. If unit is to be stored for long periods of time, drain water from unit and hose.
· Responsibility for operating and maintaining Medical Devices belongs to the user.
· Store the system in the place where the temperature is at -10-60°C (14-140°F), humidity at 10-85%RH, atmospheric
pressure at 500-1060 hPa, and the system is not subject to air with dust, sulfur, or salinity.
· When trouble is found, contact your to dealer.
· Do not insert or remove power cord with wet hands, electric shock may occur.
· Be sure not to expose the unit to water. It may result in a short or electric shock.
· Do not use this device near combustibles, or in a place where explosion may occur.
· Do not disassemble, there are no user serviceable parts inside control unit.
· Do not hit / drop the unit or handpiece. Place unit in a stable place.
· Use the fuse of specified rating. [AC120V: T1.6A 250V (Ref No.FU100, FU101), AC230V: T800mA 250V (Ref No.
FU100, FU101)]
W A R N I N G
· NuTorque
®
Lite system can be used with a hose connector that is manufactured in accordance with ISO 9168, type 2
or type 3.
· Connect to only 4-hole, 5-hole or 6-pin tubing.
· NuTorque Lite system is to be used by qualified personnel for dental treatment only, and only in a dental office.
· Patient safety must have first priority.
· Operate before use to check for vibration, noise, or overheating. If any abnormalities occur, stop using immediately
and contact your dealer.
· Grasp cord by plug to remove from outlet. DO NOT pull or yank on the cord itself.
· Do not connect / disconnect the handpiece during operation.
· Verify that the Speed Control Switch is adjusted within the allowable speed before use.
· When you install Control Box or Motor, DO NOT bend or twist the tubing or the cord. No unnatural force is needed to
do the installation.
· Never use the AC adapter for anything other than the control box of this product.
· Air Requirements: dry, free from contamination and oil. Use a compressor with a dry air system. Install an air filter if
necessary. Blow out the lines before installation.
· Do not Autoclave (or any other high temperature Sterilization) Control Box, AC Adapter, Motor Cord.
· The system functions normally in the environment where the temperature is at 0-40°C (32-104°F), humidity at 10-85%
RH, atmospheric pressure at 700-1060hPa, and no moisture condensation in the Control Unit. Use at outside of these
limits may cause malfunction.
· If the handpiece has not been used for a long period, check for noise, vibration, and overheating before use.
· The user shall be responsible for operation, maintenance and operation.
· NuTorque Lite needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information.
· Portable and mobile RF communications equipment can affect NuTorque Lite. Do not use RF equipment outskirts for
the product.
· The use of ACCESSORIES, transducers and cables other than those specified, with the replacement parts for internal
components,may result in increased EMISSIONS or decreased IMMUNITY of NuTorque Lite.
· NuTorque Lite should not be used adjacent to or stacked with other equipment.
. For safety, install the Control Unit in a place where the AC Adaptor can be easily removed. (It is possible to disconnect
the control unit from the power source by removing the AC Adaptor.)
. When installing the product, provide space of approximately 10cm around the Unit for easy access to the inlet and the
AC Adaptor.
. Be sure to use coolant air. Surface temperature will be more than 60°C if coolant air is not used.
2