background image

 

 

C A U T I O N

 

 

N O T I C E

· During vibration, the motor and the motor cord may affect computer and LAN cable. Noise could be heard during 

operation near a radio receiver.

· Turn off the Power Switch after each use. If unit is to be stored for long periods of time, drain water from unit and hose.
· Responsibility for operating and maintaining Medical Devices belongs to the user.
· Store the system in the place where the temperature is at -10-60°C (14-140°F), humidity at 10-85%RH, atmospheric 

pressure at 500-1060 hPa, and the system is not subject to air with dust, sulfur, or salinity.

· When trouble is found, contact your to dealer.

· Do not insert or remove power cord with wet hands, electric shock may occur.
· Be sure not to expose the unit to water. It may result in a short or electric shock.
· Do not use this device near combustibles, or in a place where explosion may occur. 
· Do not disassemble, there are no user serviceable parts inside control unit.
· Do not hit / drop the unit or handpiece. Place unit in a stable place.
· Use the fuse of specified rating. [AC120V: T1.6A 250V (Ref No.FU100, FU101), AC230V: T800mA 250V (Ref No. 

FU100, FU101)] 

 

 

W A R N I N G

· NuTorque

®

 Lite system can be used with a hose connector that is manufactured in accordance with ISO 9168, type 2 

or type 3.

· Connect to only 4-hole, 5-hole or 6-pin tubing.
· NuTorque Lite system is to be used by qualified personnel for dental treatment only, and only in a dental office. 
· Patient safety must have first priority.
· Operate before use to check for vibration, noise, or overheating. If any abnormalities occur, stop using immediately 

and contact your dealer.

· Grasp cord by plug to remove from outlet. DO NOT pull or yank on the cord itself.
· Do not connect / disconnect the handpiece during operation.
· Verify that the Speed Control Switch is adjusted within the allowable speed before use.
· When you install Control Box or Motor, DO NOT bend or twist the tubing or the cord. No unnatural force is needed to 

do the installation.

·  Never use the AC adapter for anything other than the control box of this product.
· Air Requirements: dry, free from contamination and oil. Use a compressor with a dry air system. Install an air filter if 

necessary. Blow out the lines before installation.

· Do not Autoclave (or any other high temperature Sterilization) Control Box, AC Adapter, Motor Cord.
· The system functions normally in the environment where the temperature is at 0-40°C (32-104°F), humidity at 10-85% 

RH, atmospheric pressure at 700-1060hPa, and no moisture condensation in the Control Unit. Use at outside of these 

limits may cause malfunction.

· If the handpiece has not been used for a long period, check for noise, vibration, and overheating before use.
· The user shall be responsible for operation, maintenance and operation.
· NuTorque Lite needs special precautions regarding EMC and needs to be installed and put into service according to 

the EMC information.

· Portable and mobile RF communications equipment can affect NuTorque Lite. Do not use RF equipment outskirts for 

the product.

· The use of ACCESSORIES, transducers and cables other than those specified, with the replacement parts for internal 

components,may result in increased EMISSIONS or decreased IMMUNITY of NuTorque Lite.

· NuTorque Lite should not be used adjacent to or stacked with other equipment.

. For safety, install the Control Unit in a place where the AC Adaptor can be easily removed. (It is possible to disconnect 

the control unit from the power source by removing the AC Adaptor.)

. When installing the product, provide space of approximately 10cm around the Unit for easy access to the inlet and the 

AC Adaptor.

. Be sure to use coolant air. Surface temperature will be more than 60°C if coolant air is not used.

2

Summary of Contents for NuTorque Lite

Page 1: ...plains an instruction that should be observed for safety reasons Read this manual thoroughly before use and operate the device properly These instructions will show you how to operate the product safety and prevent danger to you or others They are classified by the degree and or severity of danger All contents relating to safety should be oberserved Cautions for handling and operation Please Read ...

Page 2: ...the Speed Control Switch is adjusted within the allowable speed before use When you install Control Box or Motor DO NOT bend or twist the tubing or the cord No unnatural force is needed to do the installation Never use the AC adapter for anything other than the control box of this product Air Requirements dry free from contamination and oil Use a compressor with a dry air system Install an air fil...

Page 3: ...Model 265290 120V and 265291 230V Rated input AC120V 50 60Hz 41VA AC230V 50 60Hz 41VA Output AC28V 1 3A Fuse AC120V FU100 T1 6A 250V FU101 T1 6A 250V AC230V FU100 T800mA 250V FU101 T800mA 250V Dimensions W100 x D178 x H64 mm Cord Length Unshielded App 5 0m AC Adaptor Handpiece Tubing Connector Handpiece Tubing Connector Power Switch Power Switch Power Switch AC Adapter Connector AC Adapter Connect...

Page 4: ...e Motor Insert the Motor Cord Connector to the control box motor cord connector and tighten Fig 2 3 Connecting the AC adapter Insert the AC adapter plug into the AC adapter connector before plugging into wall outlet Fig 3 Never use the AC Adapter for anything other than the Control Box of this product 4 Connecting the Power Cord Insert the Power Cord into the Inlet of AC Adapter Fig 4 Plug cord in...

Page 5: ...the Power Switch When you turn on the Power Switch lamps and speed indicator will be illuminated 2 Before using this system make sure to perform Foot Air Calibration function Refer to 6 1 Foot Air Calibration Power Switch Symbol Mark Function OFF ON 3 Push the Gear Ratio Select Key to indicate the speed of the attachment you will use Fig 7 Handpiece GearRatio Indicator Speed min 1 rpm 16 1 100 2 5...

Page 6: ...wer limit of the air pressure to enable this function Perform this adjustment function after your purchase Once it is adjusted it will be memorized and there is no need to do this everytime NOTICE The displayed air pressure is only a guideline Measure the correct air pressure with an air gage The unit of displayed air pressure is kgf cm2 For other units refer below ex 0 1kgf cm2 0 01MPa 1 42psi 1 ...

Page 7: ...table between 4 Press key until it beeps more than 3 seconds to set the brightness 5 Press key and key at the same time for more than 3 seconds once again When it beeps this setting is complete 3 Program Function It is possible to program two 2 presets Speed Gear Ratio Forward Reverse Direction Once set you can choose between presets by pressing the or the key To set the values for Speed Gear Rati...

Page 8: ...t from motor 5 Disconnect motor from tubing by unscrewing coupler counter clockwise and pulling motor in the axial direction direction of E connector 6 Clean all external surfaces of attachment and motor using a small toothbrush and soft cloth with isopropyl alcohol 7 Dry thoroughly lubricate attachment 8 Sterilize Note To Prevent Damage To the Bur Use an Autoclavable Bur Block Attachments and bur...

Page 9: ...d mount the new O Rings into the groove CAUTION Effects of deteriorated O Rings Air Water leak Air Water does not turn off Vibration Difficult to connect disconnect the handpiece 2 Replacing Bulb 1 Turn off the power 2 Loosen Motor cover and detach it from the motor Remove the old light bulb using a small screw driver Fig 12 13 3 Align and insert the connector of the new bulb into the socket hole ...

Page 10: ...setting values are different from what used to be when turned off Turned off the motor while rotating Turn off after the Motor has stopped 9 Error Code Error Code Trouble Cause Check Remedy Overcurrent error In circuit Detected overcurrent in the circuit Contact your dealer Overcurrent error In software Detected overcurrent in the software Contact your dealer Retry Error Motor stopped by overload ...

Page 11: ...orque Lite uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emmissions CISPR11 class B The NuTorque Lite is suitable for use in all establishments including domestic estabilishments and those directly connected to the public low voltage power supply network that supply network that s...

Page 12: ...a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobiles radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the locatio...

Page 13: ...l information call 1 866 DTE INFO or visit our web site www dentalez com StarDental Division of DentalEZ Inc 1816 Colonial Village Lane Lancaster PA 17601 U S A Printed in Japan 2019 DentalEZ Inc Anyone wishing to return any DentalEZ Inc product to a DentalEZ facility MUST obtain the proper RETURN AUTHORIZATION NUMBER from the Customer Service Department The RA Number must appear clearly marked on...

Reviews: