1x
1x
1x
>3s
Press and hold the MENU
key (> 3 sec) until CAL ap-
pears. The display reads 'off'.
'CAL'
blinks.
Press either the UP or DOWN
key to activate the CAL
mode. The display changes
from 'off' to 'on'.
'CAL'
continues to blink.
Press the ENTER key. The
display now shows the set
pipetting volume.
'CAL'
blinks.
Press the ENTER key.
The display shows the tested
and corrected volume. The
CAL symbol is continously
displayed to confirm that an
adjustment has been made.
Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested.
'CAL'
blinks.
The instrument should be set to either the nominal volume (for example
100 µl for a 100 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 8 and 12 for procedures.
E.g., volume according to testing of volume 101,3 µl.
EASY CALIBRATION
Adjustment of the ErgoOne
®
E -8/-12
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
The calibration mode 'CAL'
Adjustment
What to do
How to do it
Keys to press
Display readout
24
· ErgoOne E -8/-12 · User Manual