Quick Programming Guide

 continued

Model Vr

3.  HOlD TIME 

(bypass this step by pressing reset)

This setting determines the elapsed time (in seconds) from the moment 

the control unit senses the patient’s weight on the Sensormat pad until 

monitoring begins.

Use the

 

    

 

button to change this time from 1-9 seconds.

Press RESET to move to Step 4.

4.  VOluME cONTROl 

(bypass this step by pressing reset)

This setting controls the alarm volume. Use the

 

    

 

button to adjust volume from 0-9.

1 is low volume, and 9 is maximum volume. 0 will mute the control unit, but the alarm 

will still trigger the nurse call system (if it is in use).

Press RESET to exit the Programming Mode

.

You have completed programming of the Model Vr.

 

The only remaining setting is Delay Time, which may be se-

lected at any time without entering the Programming Mode required for other settings.

cHANGE DElAY TIME

 

(time From weight oFF sensormat pad until alarm)

 

The number displayed indicates the number of seconds. Change the delay time as desired (1-9 seconds; 3 is typi-

cal) using the 

 

   

 

button.

To ensure the Model Vr retains your setting changes, unplug the power cord (either end) for 2-3 seconds  

and plug it back in. This locks in your settings, which will be retained even after power interruptions.

Your Bed-Check Model Vr is now ready to monitor your fall risk individual 

as soon as the Sensormat pad is properly placed and connected, and the 

nurse call system is connected (if used).