background image

6

EXTRA LOW NOISE MEASUREMENTS

When making extremely low  noise
measurements, it is a good practice to
connect the grounding plug of the SR554
to a ground point near the experiment. If a
good ground is not available near the
experiment, connect a wire from the lock-
in chassis (using a lug under one of the
chassis screws) to the grounding lug of
the SR554.

NOISE FIGURE

The noise figure describes the noise
contribution of an amplifier in a
measurement when compared to an ideal
amplifier.

The expression:

where 

N

 is the measured noise, 

A

 is the

pre-amp gain and 

e

n

rs

 is the Johnson

noise of the source impedance, describes
the noise figure contours shown below.
The optimum operating frequency can be
determined from this graph.

USING THE SR554 WITH SRS LOCK-
INS

The SR554 is not sensed through the 9
pin cable by SRS lock-in amplifiers.
Therefore the lock-in does NOT
compensate for the gain of the preamp.
Measurements made using the preamp
must be divided by the gain of the SR554.
The actual gain can be obtained from the
amplitude response curves on page 4.

USING THE SR554 WITH THE
SR810/830/850

For typical measurements the lock-in input
should be set to AC coupled, with the
shield grounded.  For low frequency
measurements (<1Hz), set the lock-in to
DC coupled, with the shield grounded,
since the SR554 can sense signals below
the lock-in’s AC coupling frequency
(0.16 Hz).

USING THE SR554 WITH THE
SR510/530

The SR510/530 is AC coupled from 0.5
Hz to 100 kHz.  Measurements below 0.5
Hz are not recommended with the
SR510/530.

Summary of Contents for SR554

Page 1: ...nsformer Preamplifier 1290 D Reamwood Avenue Sunnyvale California 94089 Phone 408 744 9040 Fax 408 744 9049 email info thinkSRS com www thinkSRS com Copyright 1999 by SRS Inc All Rights Reserved Revis...

Page 2: ......

Page 3: ...er and Buffer 500 nominal Transformer only 100 nominal See Amplitude Frequency Response Curve pg 6 Gain Accuracy 5 with fixed source resistance Gain Stability 100 ppm C Outputs Single ended or differe...

Page 4: ...2...

Page 5: ...If a longer cable is required any standard 9 pin cable will suffice since all connections are straight through When the lock in power is on the POWER indicator on the SR554 will light The SR554 output...

Page 6: ...lock in amplifier s overload indicator The SR554 s overload indicator is only functional when the unit is in the buffered mode If the unit is operated in the bypassed mode and power is applied the ov...

Page 7: ...l become 500 k If the instrument that the SR554 is connected to has an input impedance of 1 M 1 3 of the signal is lost Any significant cable capacitance will create a low pass filter with this output...

Page 8: ...e optimum operating frequency can be determined from this graph USING THE SR554 WITH SRS LOCK INS The SR554 is not sensed through the 9 pin cable by SRS lock in amplifiers Therefore the lock in does N...

Page 9: ...the lock in amplifier s input by 40 dB 100x over most frequencies DC to 500 MHz To minimize radiated noise a thick low impedance wire should be connected from the ground plug to a quiet ground point I...

Page 10: ...8...

Page 11: ...is places the lock in in a known state The power LED of the SR554 should light 3 Set the input to DC coupled grounded shield Set the time constant to 300 ms 1 s for 1 Hz measurement 12 dB oct filter a...

Page 12: ...10...

Page 13: ...3 120 BNC Connector BNC J 104 1 00003 120 BNC Connector BNC J 104 1 00014 160 9 PIN D Connector D Sub Right Angle PC Female J 109 1 00229 102 BINDING Binding Post L 101 6 00174 630 6611 TYPE 43 Ferrit...

Page 14: ...Screw Flathead Phillips Z 0 0 00209 021 4 40X3 8PP Screw Panhead Phillips Z 0 0 00221 000 SR440FOOT Hardware Misc Z 0 0 00263 052 3 22 Wire 22 UL1007 Z 0 0 00266 052 8 1 2 22 BLK Wire 22 UL1007 Z 0 0...

Reviews: