3.4 Commands
3 – 11
SCAR RANGE10
Example:
Scan Beginning Voltage
SCAB(?)
{
f
}
Set (query) the scan beginning voltage
{
to
f
}
, in volts. The default
value is
SCAB
0.0
.
The value for
SCAB
must be within the valid range for the selected
scan range,
SCAR
. Limits are the same as for the
VOLT
command.
The
SCAB
command is equivalent to the
entry field under SCAN
Setup.
SCAB 0.125
Example:
Scan Ending Voltage
SCAE(?)
{
f
}
Set (query) the scan ending
{
to
f
}
, in volts. The default value is
SCAE
0.0
.
The value for
SCAE
must be within the valid range for the selected
scan range,
SCAR
. Limits are the same as for the
VOLT
command.
The
SCAE
command is equivalent to the
entry field under SCAN
Setup.
SCAE 0.750
Example:
Scan Time
SCAT(?)
{
f
}
Set (query) the scan duration
{
to
f
}
, in seconds. The default value is
SCAT
1.0
.
The value for
SCAT
must be between
0.1
and
9999.9
. Values are
rounded to the nearest 0.1 s.
The duration time set by
SCAT
is defined as the time to scan from
the beginning voltage to the ending voltage; if the scan shape is
up-down, then the total scan period will be twice the value of
SCAT
.
The
SCAT
command is equivalent to the
entry field under SCAN
Setup.
SCAT 3600
Example:
DC205 Precision DC Voltage Source