background image

                                                         

 

 17

The image shift 
during focusing 

The nosepiece is not in the right position 

Turn it to the 
“ clicked ”position 

The image is a little 
yellow 

Not use the blue color filter 

Use the blue filter 

 
The brightness is 
not enough 
 
 

The size of the aperture diaphragm is too small 

Adjust again 

The position of the condenser is too low 

Adjust the position 

There are stains on the lens

including condenser, 

objective, eyepiece and collector

 

Clean it up 

 

2. Mechanical system 

TROUBLE CAUSE 

SOLUTION 

The image can not 
focus when using high 
magnification 
objective 

The specimen is placed inversely 
The coverslip is too thick 

Turn inversely 
Use the standard coverslip

0.17 

mm

 

The objective touch 
the specimen when 
changed from low 
magnification to the 
higher magnification 

The specimen is placed inversely 
The coverslip is too thick 

Turn inversely 
Use the standard coverslip

0.17 

mm

 

The specimen  is not 
easy to move 

The specimen holder is not fixed 

Fix it   

The binocular image is 
not coincident 

The interpupillar distance is not correct

Adjust it 

Eyes are too tired 

No diopter adjustment   

Adjust the diopter correctly 

The brightness is not suitable 

Adjust the voltage of the lamp 

     

3. Electrical system 

TROUBLE CAUSE 

SOLUTION 

The lamp can’t light 
when the switch is 
turned on

 

No power 

Check the connection of the   
power cord   

The bulb is not inserted 

Insert it correctly 

The bulb burns out 

Replace it 

The lamp burns out 
suddenly

 

Use a substandard lamp 
The voltage is too high 

Use the specified lamp to replace, 
if the problem is not solved, 
contact with the service department 

The brightness is not 
enough 

Use a substandard lamp 
The voltage is too low 

Use the specified lamp 
increase the voltage 

Summary of Contents for STM-2030BD

Page 1: ...gital biological microscope To insure safety and obtain optimum performance and familiarize yourself fully with the use of this microscope We recommend that you read the manual thoroughly before opera...

Page 2: ...er Notice 2 1 Components 4 2 Assembly 5 2 1 Assembly Diagram 5 2 2 Assembly Steps 6 3 Adjustment Operation 10 3 1 Adjustment Sets 10 3 2 Operation 11 4 Specification Table 14 5 Outfit 15 6 Troubleshoo...

Page 3: ...it will bring a serious damage to the instrument 8 Use the factory supplied power cord please II Maintenance 1 All the lenses have been well checked and adjusted It is forbidden to disassemble them y...

Page 4: ...re 8 Do not disassemble any parts of the microscope which will affect the function or decline the performance of the microscope 9 Place the instrument in a cool dry position When not using the microsc...

Page 5: ...4 1 Components STM 2030BD Digital Biological Microscope Eyepiece Digital viewing head Nosepiece Objectives Main body Condenser Collector Stage Ruler...

Page 6: ...s The number in the figure shows the assembly steps Before installing be sure every components is clean do not score any parts or glass surface Keep well with hexagon wrench provided When replacing th...

Page 7: ...ieces into the eyepiece tube until they are against each other as shown in Fig 4 Fig 4 Fig 3 NOTE Operation Conditions 1 Temperature 0 40 Maximum Relative Humidity 85 2 High Temperature High Temperatu...

Page 8: ...Clean the objective regularly for lens is susceptible to dust When operating use 10 magnification objective to search and focus specimen firstly then replace with higher magnification objective if ne...

Page 9: ...Plug the power cord into the power supply receptacle safely Make sure the connection is well Do use the supplied power cord all the time If lost or damaged select the same standard cord please 2 2 6...

Page 10: ...halogen Lamp 6V20W Replace the lamp as follows if necessary 1 Please set the main switch to O off state before replacing and make sure the bulb the lamp room and periphery are all cool enough to carry...

Page 11: ...15 Right Coarse Focus Knob Right Fine Focus Knob Tension Adjustment Collar Fig 14 Condenser Adjustment Knob Longitudinal Adjustment Knob Transversal Adjustment Knob Locking Set Left Fine Focus Knob Le...

Page 12: ...f you finish the observation with the short working distance objective and want to change another one be careful of not letting the objective touch the specimen 3 3 3 Focusing the specimen 18 19 1 Foc...

Page 13: ...r has been adjusted too It also needn t any user s operation 3 3 5 Aperture Iris Diaphragm Adjustment Fig 20 21 Turn the aperture iris diaphragm lever to adjust the aperture iris diaphragm Generally s...

Page 14: ...3 3 6 Adjusting the Interpupillary Distance Fig 22 The interpupillary distance range 55mm 75mm While looking through the eyepieces move both sliders to adjust interpupillary distance until the left an...

Page 15: ...oving Range 30mm Fine Division 0 004mm Condenser Abbe Condenser NA 1 2 Stage Double Layers Mechanical Stage 140mm 140mm Moving Range 74 50mm Illumination 6V20W Halogen Lamp 4 2 Eyepieces and Objective...

Page 16: ...nal Condenser Abbe Condenser NA 1 2 1 Optional Nosepiece Quadruple 1 Optional Illumination 6V20W Halogen Lamp 1 Optional Spare lamp 6V20W Halogen lamp 2 Optional Spare fuse 2A 2 Optional Eyepieces 10...

Page 17: ...oo thick or too thin Use the standard coverslip 0 17mm The specimen is placed inversely Reversal it back There was oil on the dry objective easily happened in 40X objective Clean it up There are stain...

Page 18: ...er magnification The specimen is placed inversely The coverslip is too thick Turn inversely Use the standard coverslip 0 17 mm The specimen is not easy to move The specimen holder is not fixed Fix it...

Page 19: ...18 The bulb flickers or the brightness is vertiginous The bulb is going to burn out Replace it The bulb is not entirely inserted into the holder Check and insert it again...

Reviews: