Protege IPG Clinician’s Manual
9
Ultrasonic scanning equipment.
The use of ultrasonic scanning equipment may
cause mechanical damage to an implanted neurostimulation system if used
directly over the implanted system.
External defi brillators.
The safety of discharge of an external defi brillator on
patients with implanted neurostimulation systems has not been established.
Therapeutic radiation.
Therapeutic radiation may damage the electronic circuitry
of an implanted neurostimulation system, although no testing has been done and
no defi nite information on radiation effects is available. Sources of therapeutic
radiation include therapeutic X rays, cobalt machines, and linear accelerators. If
radiation therapy is required, the area over the implanted IPG should be shielded
with lead.
Home and Occupational Environments
Electromagnetic interference (EMI).
Certain commercial electrical equipment
(for example, arc welders, induction furnaces, and resistance welders),
communication equipment (for example, microwave transmitters, linear
power amplifi ers, and high power amateur transmitters), and high voltage