BUTTON FUNCTIONS 

 

 
 
 
 
 
 

 

 

The “

ZERO

” button is used to turn the scale on and to zero the system before 

placing a patient onto the scale system.  When pressed, the display message will 
indicate “

CONTINUOUS WEIGH

” “

WEIGHING

” “

0.0 LB

”. The 

“ZERO” 

button will clear the memory in 

“CONTINUOUS WEIGH” 

or “

AUTO HOLD” 

mode. 

   

The “

READ

” button wakes up the display after de-activation and displays weight 

data.  (Auto power down activates after approximately 120 seconds.) 

 

The “

HOLD

” freezes the displayed weight and stores it temporarily away in 

memory.  To recall the last weight reading after the system has shut down, press 
the “

HOLD

” button before pressing any other button.  The “

ZERO

” button will 

 

clear the held memory.  

 

The “

MODE

” button allows weight data to be displayed in pounds/decimal, 

pounds/ounces, kilograms, or grams.   The display mode automatically defaults to 
the last mode used unless manually changed.   

  

 

Note: 

  Buttons are located below button names on scale base.   

 

BASIC SYSTEM OPERATION 

SETTING SYSTEM ZERO 

Ensure that the scale is free and clear of any obstructions and press the “

ZERO

” 

button.  The display message will indicate “

HANDS OFF

” “

PLEASE WAIT

” 

WEIGHING 0.00 LB

”.  The infant may now be placed on the scale for the 

weighing process.  It is recommended that the system be zeroed before each weight.     

Note

:  If there are any accessories (blanket, diaper, etc.), place them on the scale while zeroing the 

system.  This will ensure that the patient’s NET weight will be displayed.  It is recommended that 
the system be zeroed prior to each new patient.   

 

Figure 3:  Button Display 

Continued next page 

Summary of Contents for SR615

Page 1: ...0210 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SInstruments Inc by SInstrument...

Page 2: ...LY 4 REPLACEMENT PARTS AND ACCESSORIES 5 SYSTEM DESCRIPTION AND INTENDED USE 5 MAINTENANCE AND CLEANING 6 STORAGE AND TRANSPORTATION 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 8 BATT...

Page 3: ...Scale DESCRIPTION QUANTITY SCALE BASE 1 ea SCALE BASE COVER 1 ea CLEAR ACRYLIC CRADLE One cradle size included with scale base 1 ea One of the following choices included CRADLE CLEAR ACRYLIC OPEN END...

Page 4: ...the six 6 D cell batteries as described on battery compartment cover label and replace the Battery Compartment Cover Turn scale base right side up and place on a stable flat surface so that all four f...

Page 5: ...sentation of the infant s weight regardless of weight distribution The infant s weight is displayed on a 16 character dot matrix LCD With a push of a button weight data may be viewed in either pounds...

Page 6: ...cratched by abrasive cleaners Mild soap and water is recommended for general cleaning and disinfecting STORAGE and TRANSPORTATION If storing this equipment for periods longer than three 3 months remov...

Page 7: ...on AUTO POWER DOWN Approximately 120 seconds HOLD While active freezes display data and stores it temporarily in memory LAST WEIGHT RECALL Press HOLD button to recall last stored displayed reading AVE...

Page 8: ...t reading after the system has shut down press the HOLD button before pressing any other button The ZERO button will clear the held memory The MODE button allows weight data to be displayed in pounds...

Page 9: ...displays the patient s weight Zero will clear locked weight The minimum weight for the AUTO HOLD mode to function is 1 10 pounds To enable AUTO HOLD mode press and hold the HOLD button for approximat...

Page 10: ...og to digital converter The converter integrates the analog signal onto the integrating capacitor over a short interval The integrating capacitor is then discharged at a rate proportional to the refer...

Page 11: ...hold both CAL buttons 1 simultaneously for 10 seconds The display will read HOLD TO CAL as the right hand digit counts down to enter the CAL mode STEP 3 Attach cradle Offset figure should read betwee...

Page 12: ...librated weight per the tolerance table STEP 7 When settings are completed Press the HOLD button to SAVE your settings or press the READ button to CANCEL Both choices will EXIT CAL mode STEP 8 Re atta...

Page 13: ...s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc...

Page 14: ...Part No MAN615_110210 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2011 SInstruments...

Reviews: