Model: SR615 Infant Scale
Operating and Service Manual - S/N 3706+
Part No.: MAN615_110210
Page 8 of 14
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
www.srscales.com
email:
Copyright 2011
S
Instruments, Inc.
BUTTON FUNCTIONS
The “
ZERO
” button is used to turn the scale on and to zero the system before
placing a patient onto the scale system. When pressed, the display message will
indicate “
CONTINUOUS WEIGH
” “
WEIGHING
” “
0.0 LB
”. The
“ZERO”
button will clear the memory in
“CONTINUOUS WEIGH”
or “
AUTO HOLD”
mode.
The “
READ
” button wakes up the display after de-activation and displays weight
data. (Auto power down activates after approximately 120 seconds.)
The “
HOLD
” freezes the displayed weight and stores it temporarily away in
memory. To recall the last weight reading after the system has shut down, press
the “
HOLD
” button before pressing any other button. The “
ZERO
” button will
clear the held memory.
The “
MODE
” button allows weight data to be displayed in pounds/decimal,
pounds/ounces, kilograms, or grams. The display mode automatically defaults to
the last mode used unless manually changed.
Note:
Buttons are located below button names on scale base.
BASIC SYSTEM OPERATION
SETTING SYSTEM ZERO
Ensure that the scale is free and clear of any obstructions and press the “
ZERO
”
button. The display message will indicate “
HANDS OFF
” “
PLEASE WAIT
”
“
WEIGHING 0.00 LB
”. The infant may now be placed on the scale for the
weighing process. It is recommended that the system be zeroed before each weight.
Note
: If there are any accessories (blanket, diaper, etc.), place them on the scale while zeroing the
system. This will ensure that the patient’s NET weight will be displayed. It is recommended that
the system be zeroed prior to each new patient.
Figure 3: Button Display
Continued next page