SR Scales SR416i-KG Operating And Service Manual Download Page 9

Model SR416i-KG - Patient Floor Scale 
Operating and Service Manual 
Part No. MAN416I-KG_200611 

 

Page 9 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel: 716-693-5977     Fax: 716-693-5854    URL: 

www.srscales.com

 

email: 

[email protected]

     

  

Copyright 2020

 

S

Instruments, Inc. 

THEORY OF OPERATION 

 

SR Instruments patient weighing systems are digital scales. Strain-gauge force cells convert the force 
of an applied weight into an analog signal. This signal is amplified by an operational amplifier and 
converted to a digital signal by analog to digital converter. The digital signal is transferred to a micro-
controller where it is filtered, converted to appropriate units, and displayed on a liquid crystal display. 
Strain-gauge  force  cells  each  contain  four  strain  gauges  mounted  in  a  full  Whetstone-bridge 
configuration. These bridges convert the physical movement of the force cell, due to the applied mass 
on  the  system,  into  minute  changes  in  electrical  resistance.  These  changes  in  resistance  produce  a 
voltage difference across the Whetstone-bridge, which is amplified by the operational amplifier.  
The output of the operational amplifier is digitized by the analog to digital converter. The converter 
integrates  the  analog  signal  onto  the  integrating  capacitor  over  a  short  interval.  The  integrating 
capacitor is then discharged at a rate proportional to the reference voltage applied to the converter. The 
residual voltage on the integrating capacitor is then multiplied by a factor and again discharged at a 
rate proportional to the reference voltage. The residual voltage from this discharge is again multiplied 
by  a  factor  and  again  discharged.  The  time  taken  to  discharge  the  capacitor  is  proportional  to  the 
voltage from the operational amplifier, which is proportional to the applied load on the force cells. The 
time is stored as a binary number in the analog to digital converter and is transferred to the micro-
controller when the conversion is complete. 
The micro-controller averages and filters the digital output of the analog to digital converter, subtracts 
the value saved during the system zero operation and scales the filtered output, then displays the result 
on the liquid crystal display. The micro-controller performs a rolling average of data for continuous 
weigh and averages the data before locking in on the reading. 

 
 

Summary of Contents for SR416i-KG

Page 1: ...KG_200611 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S I...

Page 2: ...struments Inc TABLE OF CONTENTS PACKING CHECKLIST MODEL SR416I KG 3 ASSEMBLY 3 SYSTEM DESCRIPTION AND INTENDED USE 3 STORAGE AND TRANSPORTATION 4 MICROBAN ANTIMICROBIAL PROTECTION 4 CLEANING AND DISIN...

Page 3: ...power adapter into a wall outlet SYSTEM DESCRIPTION and INTENDED USE SYSTEM DESCRIPTION The SR416i KG Patient Floor Scale employs the latest in microprocessor and load cell technology to provide accu...

Page 4: ...TATION To maintain proper operation of this instrument storage and transport conditions should not vary outside the following conditions Relative Humidity Range of 0 to 85 Ambient Temperature Range of...

Page 5: ...eas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disin...

Page 6: ...ibrated range AUTO ZERO One button operation AUTO WEIGH Scale will automatically turn on and lock in on a stable weight AUTO POWER DOWN After 1 minute AVERAGING Automatic digital filter POWER SUPPLY M...

Page 7: ...play reads ZERO then release and wait until the display reads 0 0 When the patient is in position on the scale the weight will be displayed NOTE When the system is zeroed the Hold Recall memory is era...

Page 8: ...rnatively press the Zero Weigh button and wait for the display to indicate ZERO Release the Zero Weigh button and the display will read 0 0 Position the patient on the platform Weight will be displaye...

Page 9: ...by the operational amplifier The output of the operational amplifier is digitized by the analog to digital converter The converter integrates the analog signal onto the integrating capacitor over a sh...

Page 10: ...et in calibration mode Both procedures are found below To access all settings below remove the feet spacers and seven 7 screws holding the bottom cover on Press the calibration button Figure 2 and CAL...

Page 11: ...n of lb or kg with mode button sec Set Auto Power Down Timer between 15 and 900 seconds grav Set Gravitational Acceleration in m s2 Full Applied calibrated mass at full scale Half Used with three poin...

Page 12: ...cceleration from the CAL menu by pressing the Left Button when the display shows grav b The display will show only the four digits to the right of the decimal point 9 XXXX c Set the local gravitationa...

Page 13: ...t by pressing the Left Button to quit without saving changes e Press the Left Button again to save and the display will show SAV d Calibrate using the Two Point Calibration Curve zero and full scale N...

Page 14: ...ECTIVE ACTION Weight reading is much lower than expected Check that the platform is clear underneath stand on each corner to see if one corner is not weighing correctly System fails to perform correct...

Page 15: ...liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In n...

Page 16: ...rt No MAN416I KG_200611 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstrument...

Reviews: