BASIC SYSTEM OPERATION (CON’T) 

 

BUTTON FUNCTIONS 

The “

WEIGH

” button will display weight in pounds or kilograms as selected. 

The “

Held Wt

” button allows access to the last stable weight that was achieved. 

The “

ZERO

” button is used to zero the system. 

The “

POUNDS

” button allows weight data to be viewed in pounds, displayed in a resolution of 

5 pound intervals. 

The “

KILOS

” button allows weight data to be viewed in kilograms, displayed in a resolution of 

2 kilogram intervals. 

 

BATTERY REPLACEMENT 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

When battery replacement is needed, an indicator will appear on the display. 

STEP 1: 

Remove  the  center  Top  Panel  (Figure  4)  and  disconnect  the  two  (2)  cables  from  the 
Transducers. Turn the Top Panel over and remove the four (4) Display Cover screws and the 
clear plastic cover from the Display housing. Unscrew the two thumb screws on the Battery 
Cover and remove the battery cover (Figure 7)

STEP 2: 

Remove and replace all six (6) “D” cell batteries. Replace the Battery Cover and tighten the 
two thumb screws. 

STEP 3:

  Place the clear plastic cover on the housing and secure with the four (4) screws. 

NOTE:

 Tighten screws only until rubber seal starts to compress. 

STEP 4: 

Connect the two (2) cables from the Display housing to the Transducers. 

STEP 5:

  Press the “

ZERO/ON

” button on end of Transducer tube and confirm display is working. 

STEP 6:

  Set the Top Panel back in place by sliding it down over the Transducer assembly pins. 

Battery Cover 

Display Cover 

Figure 7: Display Assembly 

Summary of Contents for SRScales SRV714 Series

Page 1: ...age 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S by S Instruments I...

Page 2: ...ARTS and ACCESSORIES 6 SYSTEM DESCRIPTION and INTENDED USE 6 MAINTENANCE and CLEANING 7 STORAGE and TRANSPORTATION 7 THEORY OF OPERATION 7 SPECIFICATIONS 8 BASIC SYSTEM OPERATION 9 BATTERY REPLACEMENT...

Page 3: ...ruments com Copyright 2020 SInstruments Inc PACKING CHECKLIST Model SRV714 Large Animal Platform Scale DESCRIPTION QUANTITY TOP PANELS 47 in 119 cm x 31 75 in 81 cm 3 ea TRANSDUCERS 95 3 8 in 242 cm x...

Page 4: ...to battery holder imprinted instructions Attach the battery holder cover and tighten the thumb screws Re secure the plastic cover with the four 4 screws STEP 3 Remove the other two end panels from th...

Page 5: ...les are in line with the connectors on the transducer tubes Figure 4 STEP 8 After the transducers are plugged in carefully align the center panel with the four 4 alignment pins that are located in the...

Page 6: ...Scale derives its power from six 6 each D cell batteries Weight is displayed with a displayed resolution of 5 pounds or 2 kilograms The SRV714 Large Animal Platform Scale is equipped with a Bluetooth...

Page 7: ...to a digital signal by an analog to digital converter The digital signal is transferred to a micro controller where it is filtered converted to appropriate units and displayed on a liquid crystal dis...

Page 8: ...BLUETOOTH ON TIME 10 minutes from last button press AVERAGING Automatic digital filter POWER SUPPLY Six 6 D cell batteries CALIBRATION Calibration is traceable to NIST standards OPERATING CONDITIONS N...

Page 9: ...ots in the upper right corner of the SR App and select to Connect a device Figure 6B STEP 5 Select the SRV714 Bluetooth module from list of devices Figure 6C If the device is not listed the display ma...

Page 10: ...ry replacement is needed an indicator will appear on the display STEP 1 Remove the center Top Panel Figure 4 and disconnect the two 2 cables from the Transducers Turn the Top Panel over and remove the...

Page 11: ...the Display housing is out from under the scale assembly STEP 5 Press the calibration button S2 Figure 8 The display will read CAL STEP 6 Press button S4 to scroll through menu options until FULL is d...

Page 12: ...Display to the platform and securely tighten screws Re attach plastic cover to Display NOTE Tighten screws only until rubber seal starts to compress STEP 15 Connect the two Display cables to Transduc...

Page 13: ...on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event sha...

Page 14: ...AN714_200601 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2020 SInstruments Inc S By...

Reviews: