SR Instruments SR Scales SRV945-BT Operating And Service Manual Download Page 8

Model SRV945-BT 24 X 36 Platform Scale App-Enabled 
Operating and Service Manual 
Part No.:  MAN945-BT_220429 

 

Page 8 of 16 

 

S

Instruments, Inc.

600 Young Street, Tonawanda, NY  14150 

Tel:  716-693-5977     Fax:  716-693-5854    URL:  

www.srscales.com

 

email:  

[email protected]

        Copyright 2022

 

Instruments, Inc. 

 

BUTTON FUNCTIONS  

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

ZERO/WEIGH 

The  “

ZERO/WEIGH

”  button  is  used  to  zero  the  system  before  placing  a 

patient  onto  the  scale  system.    Ensure  that  nothing  is  in  contact  with  the 
weighing  surface  during  this  procedure.    The  display  will  read  “

0.0  LB

 

(or 

KG

).  The button also wakes up the display and shows the patient’s weight if it 

should Auto Power Down before the weighing process is done.   

 

HOLD/RECALL 

The “

HOLD/RECALL

” button freezes the displayed weight and stores it away 

in memory.  Press “

HOLD/RECALL

” to store the weight into memory. 

To recall last weight reading, press “

HOLD/RECALL

” button once scale has 

shut down. 
 

LB/KG 

Weight  data  may  be  viewed  in  either  pounds  or  kilograms.    Pressing  the 

LB/KG

”  button  allows  the  operator  to  toggle  between  the  two  readings.  

Pounds  are  displayed  in  a  resolution  of  0.2,  kilograms  are  displayed  in  a 
resolution of 0.1.   

 
 

Figure 3:  Button Display 

Summary of Contents for SR Scales SRV945-BT

Page 1: ...0429 Page 1 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Instruments Inc S by S Instru...

Page 2: ...4 SYSTEM DESCRIPTION AND INTENDED USE 5 STORAGE AND TRANSPORTATION 5 CLEANING AND DISINFECTING 6 SPECIFICATIONS 7 BUTTON FUNCTIONS 8 BASIC SYSTEM OPERATION 9 BATTERY REPLACEMENT 9 PAIRING TO BLUETOOT...

Page 3: ...Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Instruments Inc PACKING CHECKLIST Model SRV945 BT Portable Platform Scale...

Page 4: ...ments Inc ASSEMBLY To install included six 6 AA 1 5 volt alkaline batteries follow steps below STEP 1 Unlatch tab on the Battery Compartment cover on bottom of scale Figure 2 STEP 2 Insert batteries a...

Page 5: ...kilograms The SRV945 BT Platform scale is equipped with a Bluetooth wireless module By waking the scale up with the WEIGH button the wireless feature is turned on Then use the free SR Scales app Googl...

Page 6: ...on on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact s...

Page 7: ...ted range AUTO ZERO One button operation AUTO POWER DOWN Approximately 300 seconds AVERAGING Automatic digital filter POWER SUPPLY Six 6 AA 1 5 volt alkaline batteries CALIBRATION Calibration is trace...

Page 8: ...during this procedure The display will read 0 0 LB or KG The button also wakes up the display and shows the patient s weight if it should Auto Power Down before the weighing process is done HOLD RECAL...

Page 9: ...hing surface remains active If scale shuts down press the ZERO WEIGH button once to wake display and take reading HOLD RECALL The scale display is designed to automatically lock the reading on the dis...

Page 10: ...Android device by pressing the Bluetooth icon in the application and then pressing Scan for devices 3 Locate the scale Bluetooth module in the list of devices The scale s Bluetooth ID is listed on th...

Page 11: ...for remote wireless operation Use your Android device to display weight and control the SRV945 BT scale functions Figure 5 Ensure your remote display indicates CONNECTED before use The WEIGH button w...

Page 12: ...WEIGH button STEP 3 Place the first calibrated weight on the scale Wait for scale to stabilize note scale reading Remove weight STEP 4 Place second calibrated weight on scale Wait for scale to stabil...

Page 13: ...GH button to turn the scale on STEP 3 Press Lb Kg button and toggle to select lb STEP 4 Press the Calibration button on rear of display Display will read CAL STEP 5 Toggle the ZERO WEIGH button to rea...

Page 14: ...NUMBER STEP 11 When display reads SAVE press the HOLD RECALL button to cycle through Savd and return to CAL STEP 12 Calibration is completed Press the left button on rear of display to return to run m...

Page 15: ...ns or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said article...

Page 16: ...nual Part No MAN945 BT_220429 Page 16 of 16 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 S Ins...

Reviews: