TROUBLESHOOTING

 

 
 

SYMPTOM 

REASON/CORRECTIVE ACTION 

Display reads LO  

Battery needs replacing. 

Display reads Err 

The total weight exceeds the maximum capacity 
of the scale, remove weight from the surface. 

Memory will not Zero 

Memory will hold the last stored weight until it is 
replaced with another stored weight. It will never 
zero. 

Memory will not store 

Memory will not store weight if less than two (2) 
pounds or (0.9 kilograms) 

For additional information or assistance, phone our Service Hotline: 1-800-654-6360 

or e-mail:

 

[email protected]

 

 
 
 

 

CAUTION 

The  integrated  circuits  and  semiconductors  on 
the  printed  circuit  boards  may  be  damaged 
by  electrostatic  discharge  (ESD).    Be  sure  to 
use proper handling precautions at all times.

 

Summary of Contents for SR Scales SRV930

Page 1: ...NSRV930_220615 Page 1 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 SInstruments Inc S by S Instruments Inc Operating and Service Manual Model SRV930 Medium Capacity Counter Top Scale ...

Page 2: ...TENTS TABLE OF FIGURES 2 PACKING CHECKLIST MODEL SRV930 3 ASSEMBLY 3 ACCESSORIES 3 SYSTEM DESCRIPTION AND INTENDED USE 4 STORAGE AND TRANSPORTATION 4 CLEANING AND DISINFECTING 5 SPECIFICATIONS 6 BUTTON FUNCTIONS 7 BASIC SYSTEM OPERATION 8 BATTERY REPLACEMENT 9 CALIBRATION 10 TROUBLESHOOTING 11 WARRANTY 12 NOTES 13 TABLE OF FIGURES Figure 1 Battery Compartment 3 Figure 2 Display Button Functions 7 ...

Page 3: ...NG CHECKLIST If there are any missing or damaged parts please call the Service Hotline 1 800 654 6360 STEP 2 Turn the SRV930 scale over and slide open the battery compartment cover Figure 1 Insert the supplied 9 volt battery per the instructions in the BATTERY REPLACEMENT section Slide the battery compartment cover closed until it s secure Turn scale over STEP 3 Install End Caps over handrails ACC...

Page 4: ...aptor SRV930AC The animal s weight is displayed on a 4 digit LCD display INTENDED USE The SRV930 scale is designed for use as an animal weighing system for animals not exceeding 44 pounds Please read the following warnings to avoid injury to the animal or attendant STORAGE and TRANSPORTATION STORAGE If storing this equipment for periods longer than three 3 months remove the battery To maintain pro...

Page 5: ...ontact areas Use a soft cloth dampened with disinfectant or a damp disposable disinfectant cloth Cloth cannot be dripping wet Follow manufacturer s instruction on the proper use of commercially available disinfectants Disinfectant solutions with 1 sodium hypochlorite or 70 isopropyl alcohol are suitable for display user interface and other scale contact surfaces After disinfecting use a soft cloth...

Page 6: ...OLUTION 0 02 lb 0 5 oz 0 01 kg AUTO POWER DOWN Adjustable to 120 180 240 300 seconds and OFF MEMORY While active freezes display data and stores it in memory LAST WEIGHT RECALL MEMORY button recalls last stored weight AVERAGING Automatic digital filter POWER SUPPLY One 1 9 volt battery or optional AC adaptor CALIBRATION Calibrated to manufacturing specifications OPERATING CONDITIONS Normal operati...

Page 7: ...right 2022 SInstruments Inc BUTTON FUNCTIONS TARE ON OFF The TARE ON OFF button is used to turn the scale on and off to tare the weight or zero the display Holding the TARE ON OFF button for two seconds will manually shut down the scale Kg Lb The Kg Lb button is used to toggle between units of lb oz lb oz or kg The default units are set to pounds MEMORY The MEMORY button is used to recall the weig...

Page 8: ...eight until the memory is replaced with another stored memory the memory will not zero NOTE The MEMORY button will not store weight that is less than two pounds one kilogram HOLD ARROW INDICATOR The HOLD arrow at the left bottom of the LCD display will at first flash while weighing Once weight has stabilized the arrow stays solid REWEIGHING NEXT USE To reweigh the same or different patient be sure...

Page 9: ...ient to operate the scale STEP 1 While the tray is empty turn the scale over with the feet facing in the upward position STEP 2 Slide the battery compartment cover open as the arrow on cover indicates Figure 3 STEP 3 Remove and unclip the old 9 volt battery and properly dispose of STEP 4 Clip in a new 9 volt battery STEP 5 Replace the battery compartment cover STEP 6 Turn the scale over to an upri...

Page 10: ...ode STEP 4 Place a 20 kg calibrated weight on the scale platform Display will show a span count between 2000 16000 STEP 5 Press TARE ON OFF button one 1 time Display will read CAL and then read 20 00 kg STEP 6 Remove the weight Display will read 0 00 Calibration is now complete Press the TARE ON OFF button to turn scale off Figure 5 Calibration Button Display IMPORTANT CALIBRATION CHECK Qualified ...

Page 11: ...total weight exceeds the maximum capacity of the scale remove weight from the surface Memory will not Zero Memory will hold the last stored weight until it is replaced with another stored weight It will never zero Memory will not store Memory will not store weight if less than two 2 pounds or 0 9 kilograms For additional information or assistance phone our Service Hotline 1 800 654 6360 or e mail ...

Page 12: ...ions or liabilities on SR Instruments Inc s part and it neither assumes nor authorizes any other person to assume for SR Instruments Inc any other liabilities in connection with the sale of said articles In no event shall SR Instruments Inc be liable for any subsequent or special damages Any misuse improper installation or tampering shall void this warranty DAMAGED SHIPMENTS Title passes to purcha...

Page 13: ... Scale Operating and Service Manual Part No MANSRV930_220615 Page 13 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 SInstruments Inc NOTES ...

Page 14: ... Manual Part No MANSRV930_220615 Page 14 of 14 SInstruments Inc 600 Young Street Tonawanda NY 14150 Tel 716 693 5977 Fax 716 693 5854 URL www srscales com email sri srinstruments com Copyright 2022 SInstruments Inc S By SInstruments Inc Precision Technology in Perfect Balance ...

Reviews: