Model SRV715 Series
Operating and Service Manual
Part No. MAN715_190719
Page 13 of 16
S
Instruments, Inc.
,
600 Young Street, Tonawanda, NY 14150
Tel: 716-693-5977 Fax: 716-693-5854 URL:
email:
Copyright 2019
S
Instruments, Inc.
CALIBRATION (CONT’D)
STEP 5:
Press button “
S4
” to scroll through menu options until “
FULL
” is displayed. Press the
ZERO/WEIGH
button.
STEP 6:
Set the “
FULL
” value to the set of selected weights from Step 1. Use the
ZERO/WEIGH
button to select digit positions and use the
S4
button to change the value (Figure 5).
STEP 7:
When finished, the display will read “
SAVE
”. Press the
ZERO/WEIGH
button to save, or
press the
S4
button then
ZERO/WEIGH
to “
QUIT
”.
STEP 8:
Press the
S4
button to scroll to the menu option “
2 PT
” and press the
ZERO/WEIGH
button.
STEP 9:
Ensure nothing is in contact with the scale, then press the
ZERO/WEIGH
button when the
display reads “
ZERO
”.
STEP 10:
When the display reads “
FULL
”, place the weight from Step 1 on the platform and press the
ZERO/WEIGH
button.
STEP 11:
Press the
ZERO/WEIGH
button once more to “
SAVE”
. To exit without saving, press the
S4
button
to select “
QUIT
”, then press the
ZERO/WEIGH
button.
STEP 12:
When finished making adjustments, remove all weights and top platform from the scale.
STEP 13:
Replace the display cover and securely tighten screws. Re-attach display to the frame and
securely tighten screws. Place the top platform back on frame.
NOTE: TIGHTEN SCREWS ONLY UNTIL RUBBER SEAL STARTS TO COMPRESS
.
OVERTIGHENING MAY CAUSE WARPAGE AND GAPS
STEP 14:
Press the
ZERO/WEIGH
button to zero the scale. When the display indicates “
0.0
”, place a
calibrated weight on the scale again and check against the weight displayed. If the weight is not correct,
recalibrate. (If calibration cannot be accomplished, call the Service Department.)
TROUBLESHOOTING
SYMPTOM
REASON/CORRECTIVE ACTION
Weight reading is much lower than
expected.
Check that the scale is clean underneath, stand on
each of the four corners to see if one corner is not
weighing correctly.
For additional information or assistance, telephone our Service Hotline: 1-800-654-6360
CAUTION
The integrated circuits and semiconductors on
the printed circuit boards may be damaged by
electrostatic discharge (ESD). Be sure to use
proper handling precautions at all times.